SRS Labs SR640 Series Manual Download Page 9

7

INSTRUMENT OVERVIEW

FR ONT PANEL SUMMAR Y

FR ONT PANEL OPER ATION

The front panel has been designed to be nearly
self-explanatory. The effect of each keypress is
usually reflected by a change in the state of a
nearby LED.The front panel controls for each filter
channel are identical. The following discussion of
the front panel controls applies to both channel 1
and channel 2. 

POWER  BUTTON

The SR640 is turned on by depressing the POW-
ER button. When the unit is turned on, the model
number, serial number and firmware version num-
ber will appear briefly on the LED dis-
plays.

When the unit is powered off, all instru-

ment settings are stored in non-volatile memory.
Upon power up, the SR640 checks the consisten-
cy of the stored settings. If an error appears on
power up, the instrument is returned to its default
state. The default state for each filter channel is:

Cutoff Frequency:

5 kHz

Coupling:

DC

Source:

A

Invert:

Off

InputGain:

0 dB

Output Gain:

0 dB

Filter:

In

FR EQUENCY DISPLAY

The three digit LED displays indicate the cutoff fre-
quency of each filter.

The adjacent unit LEDs in-

dicate whether the displayed frequency is in Hz or
kHz.

The cutoff frequency for either channel may

be changed using the frequency increment and
decrement buttons to the right of the dis-
play.

When one of these buttons is pressed, the

displayed frequency will begin scrolling up or
down. Note, however, that the actual filter cutoff
frequency does NOT change until the ENTER but-
ton is pressed.

The ENTER LED will light to indi-

cate that the displayed frequency does not match
the actual filter cutoff frequency.

When the EN-

TER button is pressed, the ENTER LED will turn
off and the filter frequency will change to match
the displayed frequency.

If the frequency display

has been changed but it is not desired to change
the actual filter cutoff frequency, pressing the CLR/
LOCAL button will reset the display to the actual
filter frequency.

SOUR CE SELECT

There are two input BNC connectors provided for
each filter channel. When source LED 'A'  is on,
the input signal is measured between the center
conductor of the 'A' BNC connector and its
shield.

When the 'B' LED is on, the signal is

measured between the center conductor of the 'B'
BNC connector and its shield. When the 'A-B' LED
is on, the signal is taken between the center con-
ductor of the 'A' connector and the center conduc-
tor of the 'B' connector.

In all cases, the shields

of the 'A' and 'B' BNC connectors are connected to

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for SR640 Series

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...2 99 Notice The information described in this manual pertains to the SR640 dual channel low pass filter the SR645 dual channel high pass filter and the SR650 low pass high pass filter The manual refer...

Page 3: ...ith the SR640 9 Configuration Switches 9 Command Syntax 9 RS 232 Echo 9 COMMAND LIST Filter Commands 11 Status and Interface Commands 12 PROGRAMMING EXAMPLES Basic 13 FORTRAN 13 C 14 INSTRUMENT DESCRI...

Page 4: ...2 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 5: ...lever back into its normal position and insert the correct fuse into the fuse holder LINE FUSE Verify that the correct line fuse is installed before connecting the line cord For 100V 120V use a 1 Amp...

Page 6: ...4 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 7: ...0 FILTER SPECIFICATIONS Frequency Range 1Hz to 100 kHz with 3 digit resolution Type 8 pole 6 zero elliptic Rolloff 115 dB octave Passband Ripple 0 1 dB pk pk typical Stopband Attenuation 80 dB typical...

Page 8: ...6 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...

Page 9: ...channel may be changed using the frequency increment and decrement buttons to the right of the dis play When one of these buttons is pressed the displayed frequency will begin scrolling up or down Not...

Page 10: ...L button resets the display to the actual filter frequency If the SR640 has been placed in the REMOTE mode by a command from the computer interface press ing the CLR LOCAL button will return the unit...

Page 11: ...nt to the SR640 may be sent in either UPPER or lower case A command to the SR640 consists of 4 ASCII characters followed by an optional question mark character followed by a list of arguments separ at...

Page 12: ...e refer to the National Instruments manuals for complete details Once all the hardware and software has been con figured properly try using the IBIC program sup plied with the National Instruments car...

Page 13: ...The INVT command inverts the output signal with respect to the input n 1 or leaves it uninverted n 0 LOCL n The LOCL command sets the local remote mode of the instrument Allowed values are n mode 0 l...

Page 14: ...the byte returned by the SR640 when it is serial polled via the GPIB interface It may also be queried using the STB command SRE n The SRE command sets the service request en able byte to the value spe...

Page 15: ...ver the COM2 port 20 30 Open the COM2 port for 9600 8 bits no parity ignore handshake 40 OPEN COM2 9600 N 8 2 CS DS CD AS 1 50 60 Clear COM2 port 70 PRINT 1 80 Set channel 1 pregain and postgain 90 PR...

Page 16: ...format bn f14 0 write answer stop end Pr ogr am Exampl e 3 I BM PC Mi cr osof t C Nati onal I nstr uments GPI B car d Program to test SR640 communication using National Instruments GPIB card and Micro...

Page 17: ...loat answer sr640 ibinit SR640 locate device if sr640 0 error ibwrt sr640 PREG1 1 PREG2 2 15 set gain ibwrt sr640 FREQ 1 6 query frequency ibrd sr640 string 30 and get answer sscanf string f answer pr...

Page 18: ...n of Stage paramaters for the low pass filter are 1 6347 5493 2 8060 9507 2 0793 3 9850 2 095 1 9653 4 1 076 7 375 2 6776 For the high pass filter the parameters are simply related by the lowpass to h...

Page 19: ...7 provides 2K of battery backed up RAM U1 is a 64K ROM which contains the calibration information for the filter channels The calibration ROM is mapped as I O requiring flip flop U5B to switch between...

Page 20: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...