2 – 14
Remote Operation
Token Mode
TOKN(?)
{
z
}
Set (query) the Token Query mode
{
to
z
=
(
OFF 0
,
ON 1
)
}
.
If
TOKN
ON
is set, then queries to the SIM module that return to-
kens will return the text keyword; otherwise they return the decimal
integer value.
Thus, the only possible responses to the
TOKN?
query are
ON
and
0
.
On reset,
TOKN
is set to
OFF
.
TOKN OFF
Example:
Response Termination
TERM(?)
{
z
}
Set (query) the
h
term
i
sequence
{
to
z
=
(
NONE 0
,
CR 1
,
LF 2
,
CRLF 3
,
LFCR 4
)
}
. The
h
term
i
sequence is appended to all query responses
sent by the module, and is constructed of ASCII character(s) 13 (car-
riage return) and 10 (line feed). The token mnemonic gives the
sequence of characters.
At power-on,
TERM
is set to
CRLF
.
TERM?
Example:
3
SIM964
Analog Limiter