CALIBRATION (con’d)
STEP 10:
When the display reads “
FULL
”, place the weight from Step 1 on the platform and press the
DISPLAY
button.
STEP 11:
Press the
DISPLAY
button once more to save. To exit without saving, press
S4
to select
“
QUIT
”, then press the
DISPLAY
button
STEP 12:
When finished making adjustments, remove all weight from the scale.
STEP 13:
Replace the display box cover and securely tighten screws. Re-attach display to platform and
securely tighten screws.
NOTE: TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS.
STEP 14:
Press and hold the
DISPLAY
button to zero the platform. When the display indicates “
0.0
”,
place the calibrated weight on the scale again and check against the weight displayed. If the
weight is not correct, recalibrate. (If calibration cannot be accomplished, call the
Service
Department.)
TROUBLESHOOTING
SYMPTOM
REASON/CORRECTIVE ACTION
Weight reading is much lower than
expected.
Check that the platform is clean underneath, stand on
each of the four corners to see if one corner is not
weighing correctly.
For additional information or assistance, telephone our Service Hotline: 1-800-654-6360
or e-mail:
CAUTION
The integrated circuits and semiconductors
on the printed circuit boards may be
damaged by electrostatic discharge (ESD).
Be sure to use proper handling precautions at
all times.