CALIBRATION 

 

 
STEP 1:

 Select known calibrated weights, traceable to NIST. 

NOTE:  CALIBRATED  WEIGHT  SHOULD  BE  EQUAL  TO  OR  GREATER  THAN  THE 

MAXIMUM PATIENT WEIGHT, NOT TO EXCEED THE MAXIMUM CAPACITY 
OF THE SCALE. DO NOT USE UNCALIBRATED WEIGHTS.

 

STEP 2:

 Remove the two (2) screws holding the display housing to the platform 

STEP 3:

 Pull the display out from under the platform and remove the eight (8) screws from the display 

box cover pictured in Figure 3.   

STEP 4:

 Press the calibration button.  The display will read “

CAL

”. 

STEP 5:

  Press  button  “

S4

”  to  scroll  through  menu  options  until  “

FULL

”  is  displayed.   Press  the 

DISPLAY

 button. 

STEP 6: 

Set the “

FULL

” value to the selected weight from Step 1.  Use the 

DISPLAY

 button to select 

digit positions and use button 

S4

 to change the value. 

STEP 7:

 When finished, the display will read “

SAVE

”.  Press the 

DISPLAY

 button to save, or press 

button 

S4 

then 

DISPLAY

 to “

QUIT

”. 

STEP 8:

 Press 

S4

 to scroll to the menu option “

2 PT

” and press the 

DISPLAY 

button. 

 

 

 

 

 

 

 

 

Figure 4:  Calibration Buttons 

 

STEP 9:

 Ensure  nothing  is  in  contact  with  the  platform  then  press  the 

DISPLAY

  button  when  the 

display reads “

ZERO

”. 

IMPORTANT 

CALIBRATION  CHECK      Qualified  service  personnel  only  should  perform  this 
procedure.  Load cells have no user serviceable components and should not be tampered 
with  for  any  reason.    Re-calibration  is  generally  not  required,  but  should  be  verified 
periodically  to  ensure  accuracy.    The  recommendation  for  calibration  check  is  at  least 
once every 12 months, or as individual maintenance policy requires. 

 

Continued next page 

S4 

CAL 

Display 

Low er Lim it A pplied W eight U pper Lim it

99.8

100

100.2

199.6

200

200.4

249.5

250

250.5

399.2

400

400.8

499.0

500

501.0

748.5

750

751.5

998.0

1000

1002.0

1497.0

1500

1503.0

1996.0

2000

2004.0

C alib ratio n  T o leran ce C h art

Summary of Contents for SRV710

Page 1: ...nts Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2015 SInstruments Inc S by SR Instruments Inc Operating and Ser...

Page 2: ...S 2 PACKING CHECKLIST MODEL SRV710W 3 ASSEMBLY 4 REPLACEMENT PARTS AND ACCESSORIES 5 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTT...

Page 3: ...awanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2015 SInstruments Inc PACKING CHECKLIST Model SRV710W Zoological Scale System DESCRIPTION Q...

Page 4: ...Unpack the Scale Platform 1 and check parts against the PACKING CHECKLIST If there are any missing or damaged parts call the Service Hotline at 1 800 654 6360 STEP 2 Place the Platform upside down on...

Page 5: ...ntation of the subject s weight The SRV710W Zoological Scale derives its power from six 6 each AA batteries With the push of a button weight is displayed with a displayed resolution of 0 2 pounds or 0...

Page 6: ...e cleaners The waterproof seal can also be damaged We recommend mild soap and water for general cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longe...

Page 7: ...onds AUTO POWER DOWN After five 5 minutes BLUETOOTH ON TIME 5 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA cell batteries CALIBRATION Calibration is traceabl...

Page 8: ...play box on the underside of the scale Select the mode for weighing with this Toggle Button Figure 2 LB KG Mode Pushbutton Switch BASIC SYSTEM OPERATION Ensure that the scale is free and clear of any...

Page 9: ...LACING BATTERIES When battery replacement is needed an indicator will appear on the display STEP 1 Turn the platform over and remove the four screws on the Battery Cover STEP 2 Remove and replace the...

Page 10: ...ur strain gauges mounted in a full Wheatstone bridge configuration These bridges convert the physical movement of the force cell due to the applied mass on the system into minute changes in electrical...

Page 11: ...ght from Step 1 Use the DISPLAY button to select digit positions and use button S4 to change the value STEP 7 When finished the display will read SAVE Press the DISPLAY button to save or press button...

Page 12: ...ely tighten screws NOTE TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS STEP 14 Press and hold the DISPLAY button to zero the platform When the display indicates 0 0 place the calibrated weig...

Page 13: ...and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liabl...

Page 14: ...710W_150921 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2015 SInstruments Inc S By S...

Reviews: