SR Instruments SRScales SRV945 Operating And Service Manual Download Page 11

Model SRV949 24 X 60 Platform Scale 
Operating and Service Manual 
Part No.:  MAN949_170724 

 

Page 11 of 14 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

        

Copyright 

2017

 

Instruments, Inc. 

CALIBRATION cont’d 

SETTING CALIBRATION 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

NOTE

Ensure that nothing is in contact  with the  scale system during this procedure and that 

all  six  (6)  feet  are  touching  a  level  surface.    Remove  hands  from  the  system  when  noting  the 
displayed calibration results.  (Use a known calibrated weight traceable to NIST.)   
 

STEP 1

:  Remove the display cover from bottom of scale 

STEP 2

:  Press the “

ZERO/WEIGH

” button to turn the scale on. 

STEP 3

:

 

Press “

Lb/Kg

 

button and toggle to select lb. 

STEP 4

:

 

Press the left button on rear of display.  Display will read “CAL” 

STEP 5

:  Toggle the “

ZERO/WEIGH

” button to read “FULL” then press “

HOLD/RECALL

”. 

STEP 6

:  Display will read “440.0”, set this value to the value of the selected weight 

 

Press “

HOLD/RECALL

” to select a digit position 

 

Press the “

ZERO/WEIGH

” button to change the value of the digit 

 

When display reads “SAVE”, press the “

HOLD/RECALL

” button to cycle through 

“Savd” and return to “CAL” 

STEP 7

:  Toggle the “

ZERO/WEIGH

” button to read “2 Pt” then press “

HOLD/RECALL

”. 

Figure 5:  Calibration Button 

Figure 6:  Button Display 

Continued next page 

CAUTION 

The  integrated  circuits  and  semiconductors  on 
the  printed  circuit  boards  may  be  damaged 
by  electrostatic  discharge  (ESD).    Be  sure  to 
use proper handling precautions at all times.  

Summary of Contents for SRScales SRV945

Page 1: ..._170724 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 S Instruments Inc S by S Ins...

Page 2: ...IGURES 2 PACKING CHECKLIST MODEL SRV949 3 ASSEMBLY 4 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPER...

Page 3: ...onawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 S Instruments Inc PACKING CHECKLIST Model SRV949 Portable Platform Scale DESCRIPTION...

Page 4: ...srinstruments com Copyright 2017 S Instruments Inc ASSEMBLY To install included six 6 AA 1 5 volt alkaline batteries follow steps below STEP 1 Unlatch tab on the Battery Compartment cover on bottom o...

Page 5: ...latest in microprocessor and load cell technology to provide accurate and repeatable weight data Six 6 identically matched transducers are strategically placed to ensure an accurate representation of...

Page 6: ...oap and water is recommended for general cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longer than three 3 months remove the batteries To maintain...

Page 7: ...of displayed resolution for calibrated range AUTO ZERO One button operation AUTO POWER DOWN Approximately 120 seconds AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA 1 5 volt alkaline batter...

Page 8: ...is procedure The display will read 0 0 LB or KG The button also wakes up the display and shows the patient s weight if it should Auto Power Down before the weighing process is done HOLD RECALL The HOL...

Page 9: ...ace remains active If scale shuts down press the ZERO WEIGH button once to wake display and take reading HOLD RECALL The scale display is designed to automatically lock the reading on the display and...

Page 10: ...tton STEP 3 Place the first calibrated weight on the scale Wait for scale to stabilize note scale reading Remove weight STEP 4 Place second calibrated weight on scale Wait for scale to stabilize note...

Page 11: ...ZERO WEIGH button to turn the scale on STEP 3 Press Lb Kg button and toggle to select lb STEP 4 Press the left button on rear of display Display will read CAL STEP 5 Toggle the ZERO WEIGH button to re...

Page 12: ...TEP 11 When display reads SAVE press the HOLD RECALL button to cycle through Savd and return to CAL STEP 12 Calibration is completed Press the left button on rear of display to return to run mode Allo...

Page 13: ...nc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments In...

Page 14: ...art No MAN949_170724 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2017 S Instruments...

Reviews: