Model SR755KG Series Wheelchair Scale
Operating and Service Manual - S/N 1913+
Part No.: MAN755KG_100929
Page 13 of 16
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2010
S
Instruments, Inc.
TROUBLESHOOTING
SYMPTOM
REASON/CORRECTIVE ACTION
Inaccurate w
heelchair
hair and
onto the scale (see directions for
BASIC SYSTEM
OPERATION
).
eight readings
Scale system MUST be zeroed with empty w
BEFORE patient is positioned in the wheelc
The characters only appear on half of the
display.
Press the “
WEIGH
” button or remove
battery.
Wait five seconds, then re-install the battery and try the
“
WEIGH
” button again.
one
The display lights appear to work but do
ot respond to button activa
Button may not have “bounced” back up after being
and inspect buttons.
Make sure the rubber “boot” is not sitting too high. Re-
n
tion.
pressed. Remove the faceplate
attach plate.
The display shows no reading at all.
Check to make sure batteries are installed correctly (see
directions for
BATTERY REPLACEMENT
).
Check display cable to ensure it is connected securely.
For additional information or assistance, telephone the Service Hotline: 1-800-654-6360
or e-mail: