CALIBRATION (CON’T) 

 

SETTING CALIBRATION 

 

NOTE:

 Ensure that nothing is in contact with the scale system during this procedure. Remove hands from 

the system when noting the displayed calibration results. 
 

STEP 1:

 Remove the two (2) screws on the display assembly cover. Remove the display assembly from the 

scale.  Put  the  scale  system  into  the  Calibration  Mode  by  switching  the  calibration  switch  on  the  display 
board (Figure 6). 

STEP  2:

  Select  known  calibrated  weights,  traceable  to  NIST,  up  to  the  Full  Scale  value  (maximum 

capacity). 

STEP  3:

  Press  the  “

MENU”

  button  until  “

FULL  1000.0Lb

”  is  displayed.  Set  the 

FULL

  value  to  the  actual 

quantity  of  calibrated  weight  being  used  for  Full  Scale.  Use  the  “

UP”

  arrow  button  to  select  the  digit  and  the 

DOWN”

 arrow button to change digit. Press “

ENTER

” to save changes. Press “

WEIGH

” button to abort any 

changes. 

STEP 4:

 Press the 

MENU

 button until “

HALF 500.0Lb

” is displayed. 

NOTE

:  The Half-Scale value is a value between zero and the Full Scale values.  It is usually close to half 

the Full Scale value. 

Set the 

HALF

 value to the actual quantity of calibrated weight being used for Half Scale. Use the “

UP”

 

arrow button to select the digit and the “

DOWN”

 arrow button to change the digit. Press 

“ENTER”

 to save 

changes. Press “

WEIGH

” to abort the changes. 

STEP 5:

 Press 

MENU

 button until “

3 PT CAL

” is displayed. Press the “

UP”

 arrow button. 

STEP  6:   

Display  will  read  “

ZERO

”.    Ensure  that  all  weight  has  been removed  from  the  platform  and 

press the “

UP” 

arrow button. 

STEP  7:

    Display  will  read  “

ADD  HALF

”.    Place  the  calibrated  weight  on  platform  for 

HALF 

Scale.  

Allow weight to stabilize.  Press “

UP”

 arrow button to save change. 

STEP  8:

    Display  will  read  “

ADD  FULL

”.    Place  the  calibrated  weight  on  platform  for 

FULL 

Scale.  

Allow weight to stabilize.  Press “

UP” 

arrow to save change.   

STEP 9

: Press 

“ENTER” 

button to save the calibration, or “

WEIGH

” button

 

to exit without saving

.

 

STEP 10:

 Switch the scale system out of the Calibration Mode on the display board (Figure 6). 

NOTE

: Ensure that no wires have become unplugged during calibration. 

STEP 11:

 Place the display assembly into the scale frame. Attach using the two (2) display cover screws. 

Summary of Contents for SR Scales SR7020i

Page 1: ...ICROBAN_200520 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by...

Page 2: ...INSTALLATION 3 REPLACEMENT PARTS AND ACCESSORIES 6 SYSTEM DESCRIPTION AND INTENDED USE 6 MICROBAN ANTIMICROBIAL PROTECTION 6 MAINTENANCE AND CLEANING ERROR BOOKMARK NOT DEFINED STORAGE 7 SPECIFICATION...

Page 3: ...t location for the scale by considering these recommendations Have free wall hallway space to roll equipment onto the scale from either side Platform extends 42 into the hallway when folded down Hallw...

Page 4: ...ure 2 NOTE The wall mounting bracket does not need to be centered on the crossbar but must be located between the bumpers STEP 5 For stud walls use a stud finder and locate any studs in the mounting a...

Page 5: ...head screws long enough to anchor into the wall or the concrete anchors NOTE The screws need to be a flat head type screw so there is adequate clearance for the scale frame to slide into the wall bra...

Page 6: ...atform Scale system has a 38 x 58 96 5 cm x 147 3 cm platform weighing surface The scale system has a programmable Automatic Power Down APD for adjusting scale on time INTENDED USE The SR7020i Wall Mo...

Page 7: ...olutions with 1 sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth dampened with clean water and d...

Page 8: ...e button operation after platform is left down AUTO POWER DOWN Adjustable between 30 to 300 seconds AVERAGING Automatic digital filter POWER SUPPLY Medical Grade AC Power Supply CALIBRATION Calibratio...

Page 9: ...Press to recall the last stored weight The stored weight will be erased when the scale is zero or another stable weight is stored BMI Press to calculate BMI When the BMI is pressed the default value H...

Page 10: ...etting time and date BASIC SYSTEM OPERATION SETTING SYSTEM ZERO Scale will auto zero when the platform is first lowered Make sure the scale is free and clear of any obstructions When platform is left...

Page 11: ...he amplifier is configured to current sum the output of each cell with potentiometers serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiomete...

Page 12: ...DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing and holding the left ZERO button STEP 3 Place the calibrated weight on the scale Wait for scale to stabilize note scale re...

Page 13: ...abort any changes STEP 4 Press the MENU button until HALF 500 0Lb is displayed NOTE The Half Scale value is a value between zero and the Full Scale values It is usually close to half the Full Scale v...

Page 14: ...WEIGH button or disconnect power Wait five seconds then re connect the power and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure th...

Page 15: ...r liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In...

Page 16: ...No MAN7020I_MICROBAN_200520 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstr...

Reviews: