Add 10lb load. When ADC reference value is stable, press OK. This will confirm 10lb load is
calibrated. Weight point will change to 2 which corresponds 20lb.
Add Weight 20lb for calibration
•
When ADC value gets stable, press OK button. This will complete calibration for weight 20lb.
Weight point will change to 3. Add weight for 30 lb and press Ok button.
•
Continue procedures i.e. add weight and press OK button till weight of point 15 (for 150lb)
•
When calibration is done for 150lb, Press OK to check measurement or left navigation button
for going back to setting menu.
•
When Ok button is pressed, calibration procedure is completed.
•
For calibrating other cable sizes, above procedures need to be followed i.e. starting calibration
from 10 lb and ending at 150lb.
Device Info
For getting device info, press Settings button->Navigate downward to select ‘Info’ and press OK. It
displays meter and Sensor Board Firmware version as well as Battery Voltage.
Setting Sensor ID
•
Sensor ID can be set by pressing Setting button. Use Up/Down navigation key to select Sensor
ID option (it will get highlighted and arrow will point to it). Now Press OK to enter Sensor ID
settings
•
Use Left/Right Navigation key to point to the digit to be set. Selected digit will get underlined.