![Sport-Tec LOJER 4040X Manual Download Page 50](http://html1.mh-extra.com/html/sport-tec/lojer-4040x/lojer-4040x_manual_1349488050.webp)
Copyright © Lojer Oy, 2021 | 05/2021
12
3 Using the device
1.
Note!
Do not use the electrical functions of the device non-stop for longer than the permissible two
(2) minutes. Longer continuous use may cause the transformer to overheat. If you use electrical
functions non-stop for two (2) minutes, keep to the operating time ratio and do not use any electrical
functions for 18 minutes
.
Make sure that the accessories do not hit anything when adjusting/moving the table.
Make sure the device is functional before use.
WARNING!
Make s
ure that the cord doesn’t get stuck between parts of the frame or under the
castors
Examination tables are electrically adjustable by hand control, foot control or with the hands-free adjustment
bar. The table is adjusted by pushing a button from the hand/foot control. When the adjustment bar is pushed,
the table/back section lifts and when the bar is lifted, the table/back section lowers. Movement stops when the
button/bar is released. In fault situation the movement can be stopped by holding down the button for opposite
direction. With hands-free adjustment bar, the movement can be stopped by holding the bar on the opposite
direction.
3.1 Adjustment range
Adjustment range of the 4040X examination table is shown below (Figure 2).
-
Height adjustment
40-95 cm, 4040XH (hydraulic) 45-90cm
-
Back section adjustment
73°
-
Trendelenburg (option)
max. 12 °
Figure 2: Adjustment range of 4040X