UTM20 Series Ultrasonic Transit-time Flowmeters
7.6 Installing a Meter with a Remote Transmitter and
Adjustable Transducers
-
Locate the transmitter within the length of the transducer cables supplied or exchange the cable
for one of proper length.
-
Install the transducers after entering the pipe settings into the transmitter and determining the
spacing and mounting method.
-
See pages 9 and 10 for enclosure and mounting dimension details. Allow enough room for door
swing, maintenance and conduit entrances.
Important
When routing wires to the transmitter, make sure the cables are not twisted, pinched or hanging loosely.
1.
Attach the mounting bracket to a wall (with 4 customer-supplied #8 or M4 screws, see “Wall Mount
Bracket” on page 9 for dimensions) or to a pipe (with mounting straps).
2.
Align the transmitter's bracket with the mounted bracket. Use a 4 mm hex tool to secure the 4
provided screws from the sides through the mating holes. See Figure 19 .
Fig. 17
Fig. 18
3.
If necessary, you can rotate the mounting bracket in 90° increments to accommodate the final
orientation of the transmitter. From inside the enclosure, remove the 4 screws holding the bracket.
Rotate the bracket and replace the screws. See Figure 17.
Holes for Screws
for Wall Mounting
Holes for Screws
for Wall Mounting
Slots forStraps
for Pipe Mounting
Slots forStraps
for Pipe Mounting