background image

Operating Manual

12

1300-87075-1A

5.4.1  Recalling a Protocol

To recall a stored program, touch the 

Store/Recall

 button on the 

Control Panel

 screen (Figure 4). In the 

Saved 

Protocols

 screen, select the protocol to highlight the box. Then touch 

Select

 to send protocol to the 

Control 

Panel

 screen.

5.4.2  Storing a New Protocol

To store a new program, touch the 

Store/Recall 

button on the 

Control Panel

 screen (Figure 4). In the 

Saved 

Protocols

 screen, touch the 

Add 

button as shown in Figure 7. This will bring up the Keyboard screen, a simplified 

version of the standard keyboard for a computer.

 

Figure 8. 

Keyboard Screen

The 

Clear

 key deletes whatever has been entered in the label box. Touching the 

Done

 key enters the label 

shown in the box.
Touch the 

Save

 button to save the label shown in the box. The label will appear as the name of the program on 

the 

Saved Protocols 

screen (Figure 7).

The 

Cancel

 key does not change anything on the screen, but returns the display to the 

Saved Protocols

 screen.

To run the newly saved protocol, touch the protocol to highlight the box. Then touch 

Select

 to send the protocol 

to the 

Control Panel 

screen. Review the parameters and touch the 

Run

 button to initiate the protocol.

5.4.3  Edit a Protocol

To edit a saved protocol, touch the protocol name to highlight the box on the 

Saved Protocol

 screen (Figure 7), 

then touch the 

Edit

 button. From the 

Edit Protocol

 screen, change the name by touching inside the box. This will 

bring up the Keyboard screen, follow the instructions listed in the Storing a Protocol section 5.4.2. The run time, 

rate, rest time, and cycles can also be adjusted in the Edit Protocol screen.

Summary of Contents for 8200 Planetary Mill

Page 1: ...Operating Manual 1 1300 87075 1A 8200 Planetary Mill High Energy Planetary Ball Mill...

Page 2: ...oduced or transmitted in any form or by any means elec tronic or mechanical including photocopying and recording except as may be expressly permitted by the 1976 Copyright Act or in writing from the p...

Page 3: ...ed Protocols 11 6 0 Loading the Grinding Jar 13 7 0 Locking the Grinding Jar in the Clamp 14 8 0 Balancing the Jar 14 8 1 Adjusting the counterweight 14 9 0 Operation 16 9 1 Mixing and Grinding 16 9 2...

Page 4: ...Operating Manual 4 1300 87075 1A 13 0 Troubleshooting 23 14 0 Warranty 24 14 1 Product Specifications 24 14 2 To Arrange a Return Shipment 24 15 0 Instrument Disposal 24 16 0 Contact Us 25...

Page 5: ...een display sleek modern design forced air cooling a safety interlock and a choice of hardened steel or stainless steel grinding jars and media which are purchased separately The jar containing the sa...

Page 6: ...opean plug supplied Circuit Breaker Same as ON OFF switch Safety Features Door locks while running Interlock prevents mill from running if door is not closed Manual door release latch on top of the un...

Page 7: ...If there is any visible damage notify Spex and file a claim with the carrier immediately To unpack the Planetary Mill loosen and remove the eight wing nuts on the lower edge of the crate two on each...

Page 8: ...motor and timer is located at the back of the unit To turn on the power press down left the bar side of the switch Before plugging in the mill set the power to off by pressing down right the circle O...

Page 9: ...d operated through a series of touch screen displays Transition between screens and all programming and operating commands are done by touching the screen with a fingertip or stylus Do not use a sharp...

Page 10: ...the sample to dissipate the frictional heat from the grinding balls 3 Cycle 2 unit runs for another 2 minutes at a rate of 600 rpm Note If reverse was on active the planetary wheel would reverse ever...

Page 11: ...time rest time cycles and rate touch the plus or minus buttons on the Control Panel screen Figure 4 Touching the plus increases the parameter while touching the minus decreases the parameter Tap or ho...

Page 12: ...shown in the box Touch the Save button to save the label shown in the box The label will appear as the name of the program on the Saved Protocols screen Figure 7 The Cancel key does not change anythin...

Page 13: ...the parameters by using the left and right arrows If the protocol is modified MOD will appear next to the protocol name e g Shale Rock MOD 6 0 Loading the Grinding Jar Before using a planetary mill g...

Page 14: ...kwise downward unclasping it from the Y cross bar lift the locking knob assembly upward as shown in Figure 11 When placing the jar in the clamp make sure the pins are fitting in the proper holes Once...

Page 15: ...ar 500 mL 9 9 lb Stainless Steel Grinding Jar 250 mL 7 2 lb Stainless Steel Grinding Balls 40 mm 0 575 lb Stainless Steel Grinding Ball 30 mm 0 242 lb Stainless Steel Grinding Ball 20 mm 0 070 lb Stai...

Page 16: ...jar Dry grinding is the simplest approach and most often used If a sample is not being ground fine enough the user can decrease the amount of sample increase grinding time use a denser grinding mediu...

Page 17: ...remain closed during a run To open the cabinet grasp the handle and rotate counter clockwise to slide open the door to its open position Figure 16 Door Operation 9 2 1 Manual Door Release Latch To op...

Page 18: ...too warm to touch use gloves that are heat resistant The 500 mL grinding jar can be emptied by using the sieve attachment to separate the grinding balls from the ground sample Follow the steps below...

Page 19: ...rind a portion of sample B without contaminating it with sample A If a cleaned grinding jar will not be used immediately store in a plastic bag Hardened steel grinding jars should always be dried and...

Page 20: ...s Screen 10 1 Run History To recall the run history go to the Home screen and touch the Run History button The date time user run label and run protocol data are stored on this screen Figure 20 The ru...

Page 21: ...ubricated for life and require no maintenance The shaft bearings of the motor are lubricated for ten years of intermittent use or one year of heavy use The 8200 Planetary Mill is intended for intermit...

Page 22: ...its home position front center of door opening Turn off power to the unit and disconnect the electrical cord Wait a few minutes reconnect the electrical cord and turn on power to the unit If issue pe...

Page 23: ...Stop to clear display Display shows Error Mech Failure Mechanical issue Inspect the interior for loose connections Turn power OFF then ON Display reads Run Complete but door is locked Door opened too...

Page 24: ...information in this manual However changes may be made from time to time to improve performance of this product so specifications are subject to change without notice Therefore slight changes that ar...

Page 25: ...ing Manual 16 0 Contact Us Within the United States contact us at 1 732 623 0465 fax 1 732 906 2492 sampleprep antylia com or spexsampleprep com Outside the United States contact the Spex SamplePrep r...

Reviews: