10 

mL 

   4. MEASURING  PROCEDURE 

 

 

4-1 Measurement 

 

Before the measurement, you should set the 

measurement mode to " Free Chlorine " or " Total 

Chlorine ". For setting procedure, refer to chapter 

4-3, page 8. 

 

 

1)

 Power ON the meter by pressing the Power Button  once. 

The display will show " Free ( TOTAL ) " for approx. 1 second. 

then show " CAL 0 ". The meter is now ready for the ZERO 

setting procedures. 

 

2)

 Fill the measurement liquid into the testing bottle. 

 

 

 

Note :  

Be sure to fill the 

testing bottle to the 

top of the 10ml line.  

 

3)

 Open the test camber cover and insert the testing 

bottle into the test chamber. Make sure the testing 

bottle is seated fully and oriented correctly.  The 

white mark on the bottle should point towards the 

white mark on the test chamber. Close the cover. 

 

Note : 

Be sure orient the 

testing bottle into 

the test chamber 

correctly.  

 

 

 

 

 

ZERO setting for the liquid 

Summary of Contents for 860043

Page 1: ...METER marks a stepforward foryouintothefieldofprecision measurement Although this CHLORINE METER is a complex and delicate instrument its durable structure developed Please read the following instructions carefully and always keep thismanual within easy reach OPERATION MANUAL ...

Page 2: ... 2 SPECIFICATIONS 2 3 FRONT PANEL DESCRIPTION 4 4 MEASURING PROCEDURE 5 4 1 Measurement 6 4 2 Free Total Chlorine mode selection 7 4 3 Data Hold 7 4 4 Data Record Max Min reading 9 5 CALIBRATION PROCEDURE 9 6 BATTERY REPLACEMENT 10 ...

Page 3: ...on are included as standard accessories The unique optics structure enables the instrument to read with high resolution 0 01 ppm mg L Splash waterproof on the front panel Jumbo LCD easy readout Microprocessor circuit assures maximum possible accuracy provides special functions andfeatures Battery operated for field and on sitetesting convenience Data hold function for freezing the desired valueon ...

Page 4: ...adapta tion of the USEPA Method 330 5 and Standard Method 4500 Cl G Note The reaction between free total chlorine and the DPD reagent cause a pink tint in the sample Response time Less than 10 seconds Sample volume 10 mL Data Hold Freeze the display reading Memory Recall Maximum Minimum value Display Samplin g Time Approx 1 second Power off Auto shut off saves battery life or manual off by push bu...

Page 5: ...e Chlorine standard solution CF 01 1 PC 1 0 ppm Total Chlorine standard solution CT 01 1 PC Zero Chlorine standard solution CL 01 1 PC Empty testing bottle 2 PCs Clean cloth 1 PC Free Chlorine DPD powder 10 PCs Total Chlorine DPD powder 10 PCs Optional Accessories Free Chlorine DPD powder 10 PCs Model CFP 10 Total Chlorine DPD powder 10 PCs Model CTP 10 Empty testing bottle Model 0601 1 0 ppm Free...

Page 6: ... ZERO Button 3 8 REC Button MAX MIN Button 3 9 Battery Compartment Cover 3 10 Free Chlorine 1 0 ppm standard solution 3 11 Total Chlorine 1 0 ppm standard solution 3 12 Zero Chlorine standard solution 3 13 Empty testing bottle 1 3 14 Empty testing bottle 2 3 15 Clean Cloth 3 16 Free Chlorine DPD powder 10 PCs 3 17 Total Chlorine DPD powder 10 PCs ...

Page 7: ...ter is now ready for the ZERO setting procedures 2 Fill the measurement liquid into the testing bottle Note Be sure to fill the testing bottle to the top of the 10ml line 3 Open the test camber cover and insert the testing bottle into the test chamber Make sure the testing bottle is seated fully and oriented correctly The white mark on the bottle should point towards the white mark on the test cha...

Page 8: ...Chlorinemeasurement itshould use the Free Chlorine DPD powder FortheTotalChlorinemeasurement itshould use the Total Chlorine DPD powder 7 Wait approximately one minute then insert the testing bottle into the testing chamber 8 Press the Test Button once the display will flash TEST wait approximately 10 seconds then the display will show the measurement value After the testing you should wash the Te...

Page 9: ...o select the desiring function Free Chlorine or Total Chlorine then press the REC Button to confirm and save theselection mode into the memory Note Fortheswingingpoolapplication it istypicalto select the Free Chlorine function Fortheindustrialwaterpollutionapplication typical to select the Total Chlorine function 4 3 Data Hold 1 During the measurement pressing the Hold Button once will hold the me...

Page 10: ...ss the REC Button once the REC MAX symbol along with the maximum value will appear on the display Note If you want to delete the maximum value press the Hold Button once then the display will return to the REC symbol only 2 Press the REC Button again the REC MIN symbol along with the minimum value will appear on the display Note If you want to delete the maximum value press the Hold Button once th...

Page 11: ...echlorine are independent 2 Calibration requires the following two solutions Zero Chlorine standard solution 1 0 ppm standard solution Free or Total 3 1 Turn the meter on 2 Press and hold the CAL button until CAL appears on the screen 3 Let go of the CAL button The screen will display 0 00 prompting for the Zero standard solution Total chlorine Zero 1ppm chlorine calibration Free chlorine ...

Page 12: ...press the TEST CAL button once The meter will flash CAL on the screen for up to ten seconds Note If the calibration was successful the screen will stop blinking and display 0 0 indicating that it is ready to test samples 6 BATTERY REPLACEMENT 1 When the left corner of LCD display shows it is necessary to replace the battery However in spec measurements may still be made for several hours after low...

Reviews: