4. MEASURING PROCEDURE
4-1 Measurement
Before the measurement, you should set the
measurement mode to " Free Chlorine " or " Total
Chlorine ". For setting procedure, refer to chapter
4-3, page 8.
1)
Power ON the meter by pressing the Power Button once.
The display will show " Free ( TOTAL ) " for approx. 1 second.
then show " CAL 0 ". The meter is now ready for the ZERO
setting procedures.
2)
Fill the measurement liquid into the testing bottle.
Note :
Be sure to fill the
testing bottle to the
top of the 10ml line.
3)
Open the test camber cover and insert the testing
bottle into the test chamber. Make sure the testing
bottle is seated fully and oriented correctly. The
white mark on the bottle should point towards the
white mark on the test chamber. Close the cover.
Note :
Be sure orient the
testing bottle into
the test chamber
correctly.
ZERO setting for the liquid