Page 1: ...1 Anemometer SD Card Datalogger 850023 Instruction Manual...

Page 2: ...chanical or other means that do not yet exist or may be developed including photocopying recording or any information storage and retrieval system without the express permission from Sper Scientific A...

Page 3: ...URES 5 MATERIALS SUPPLIED 6 POWER SUPPLY 7 METER COMPONENTS 8 KEYPAD 10 SETUP MODE 11 MEASUREMENT PROCEDURES 15 DATALOGGER 19 BATTERY REPLACEMENT 22 TROUBLESHOOTING 23 PC CONNECTION 27 OPTIONAL ACCESS...

Page 4: ...ermometer This meter features a real time SD memory card datalogger Standard portable SD memory cards provide unlimited data storage and upload pre formatted data directly to Excel eliminating the nee...

Page 5: ...datalogger Manual datalogger option Fast response time for air temperature Highly accurate microcomputer circuit RS232 or USB PC connection Direct upload of data to Excel Detachable probe for measurin...

Page 6: ...6 MATERIALS SUPPLIED Meter Probe SD Card 6 AA Batteries Instruction Manual Soft Carrying Case...

Page 7: ...ptional 9 Volt DC adapter See page 22 for battery replacement instructions Plug the adaptor into the power port labeled DC 9V located on the side of the meter Note When using the adapter to run the me...

Page 8: ...8 METER COMPONENTS Top of Meter Bottom of Meter Side of Meter Anemometer Probe Input Socket Type K J Thermometer Socket SD Card Socket RESET RS232 Output DC 9V Adapter Input Socket...

Page 9: ...9 METER COMPONENTS Back of Meter Probe Tripod Mounting Screw Battery Cover Screws Battery Compartment Cover Stand Probe Head Probe Handle Probe Plug...

Page 10: ...10 KEYPAD LCD Display POWER ESC Backlight SET TIME CHECK LOGGER SAMPLING TIME CHECK HOLD NEXT FUNCTION REC ENTER UNIT...

Page 11: ...but not while utilizing the datalogger function Once selections are saved the meter will default to the selected preferences when the meter is turned off and on again Once a selection is saved the met...

Page 12: ...our minute and second Note This procedure adjusts the meter s internal clock The internal clock will function when the meter is turned off but only when the batteries have adequate power not with low...

Page 13: ...o auto power off disabled Press ENTER to save the selection Touch Tone 1 Enter the touch tone function from Setup Mode see page 11 bEEP appears on the LCD 2 Press or to select yes touch tone enabled o...

Page 14: ...onds Press ENTER to save the value SD Memory Card Format Enabling this function will format the SD memory card to work specifically with your meter Formatting the SD card will erase any previous memor...

Page 15: ...ired function Air velocity temperature An appears on the LCD Type K J thermometer tP appears on the LCD Note The meter will default to the last function setting used when it is turned off and on again...

Page 16: ...on the upper display and the air temperature value appears on the lower display Note Point the yellow dot indicator on the probe head toward the air flow Type K J Thermometer 1 Press POWER to turn th...

Page 17: ...ppear on the LCD 2 Press HOLD again to release the hold function HOLD will disappear from the LCD Maximum and Minimum To record maximum and minimum readings 1 When measuring any parameter press REC to...

Page 18: ...atically when the meter is turned on 2 Press to turn the backlight off 3 Press to turn the backlight on View Real Time Clock To view the time function during normal measurement not during datalogging...

Page 19: ...alogging 1 Set the sampling time to 1 second Refer to page 14 2 Press REC REC will appear on the LCD 3 Press LOGGER REC will flash on the LCD and the alarm will sound while the measurement data and ti...

Page 20: ...also be recorded on the SD card Note To enable disable the touch tone feature see page 13 To change the position number press The position number will flash on the LCD Press or to set the position num...

Page 21: ...n the data is saved to the AMA01001 XLS file until the data reach 30 000 data columns A new file will then be generated i e AMA01002 XLS 3 The folder AMA01 will hold 99 files A new route i e AMA02 wil...

Page 22: ...ve the old batteries and replace with six new AA batteries ensuring correct polarity 4 Replace the battery cover Tighten the screws on the battery cover to secure to the meter TROUBLESHOOTING System R...

Page 23: ...s SD card slot if built into the computer or into a SD card adapter ensuring that the adapter is connected to the computer 3 Turn the computer on and run the Excel program 4 Download the saved data fi...

Page 24: ...This meter has a RS232 PC serial interface via a 3 5 mm terminal The data output is a 16 digit stream that can be utilized for the user s specific application A RS232 lead with the following connectio...

Page 25: ...0 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0 Each digit indicates the following status DO End word D1 D8 Display reading D1 LSD D8 MSD For example If the display reading is 1234 then D8 to D1 is 00001234 D9 Decima...

Page 26: ...mile h 12 F 02 Km h 10 m S 08 ft min 11 D13 When send the upper display data 1 When send the lower display data 2 D14 4 D15 Start word PC CONNECTION RS232 Format 9600 N 8 1 Baud rate 9600 Parity No pa...

Page 27: ...ECTION 800060 to 77 Type K Thermocouple Probes 840057 RS232 Computer Cable 840059 SD Card 840090 Water Resistant Instrument Pouch 840093 Field Tripod 840094 USB Computer Cable 840097 AC Adaptor 850080...

Page 28: ...lometers per hour Ft min FPM feet per minute Knots nautical miles per hour Mile h mph miles per hour Air Temperature C F Type K J Thermometer C F Temperature Compensation Automatic temperature compens...

Page 29: ...ower adapter is optional Power Current Normal Operation without use of the datalogger or backlight Approximately DC 15 mA Datalogger Operation backlight is off Approximately DC 28 mA If the backlight...

Page 30: ...g 0 2 m S Km h 1 4 to 126 0 Km h 0 1 Km h 2 reading 0 8 Km h mph Mile h 0 9 to 78 3 mph 0 1 mph 2 reading 0 4 mph Knots 0 8 to 68 0 Knots 0 1 Knots 2 reading 0 4 knots FPM Ft min 79 to 6890 FPM 1 FPM...

Page 31: ...1300 0 C 50 1 to 100 0 C 0 1 C 0 4 reading 0 5 C 0 4 reading 1 C 58 0 to 2372 0 F 58 1 to 148 0 F 0 1 F 0 4 reading 1 F 0 4 reading 1 8 F Type J 50 0 to 200 0 C 50 1 to 100 0 C 0 1 C 0 4 reading 0 5...

Page 32: ...not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal and vo...