background image

1

RH/Temperature      

 SD Card Datalogger

800021

Instruction Manual

Summary of Contents for 800021

Page 1: ...1 RH Temperature SD Card Datalogger 800021 Instruction Manual ...

Page 2: ...ot be reproduced or transmitted in any form or by any means electronic mechanical or other means that do not yet exist or may be developed including photocopying recording or any information storage and retrieval system without the express permission from Sper Scientific SPER SCIENTIFIC LTD 8281 E Evans Rd Suite 103 Scottsdale AZ 85260 480 948 4448 ...

Page 3: ...ION 4 FEATURES 5 MATERIALS SUPPLIED 6 POWER SUPPLY 6 METER COMPONENTS 7 KEYPAD 9 SETUP MODE 10 MEASUREMENT PROCEDURES 14 DATALOGGER 18 BATTERY REPLACEMENT 21 TROUBLESHOOTING 21 OPTIONAL ACCESSORIES 22 SPECIFICATIONS 23 WARRANTY 28 ...

Page 4: ...K J thermocouple thermometer This meter features a real time SD memory card datalogger Standard portable SD memory cards provide unlimited data storage and upload pre formatted data directly to Excel eliminating the need for cables or software Alternatively manually record 99 readings for on screen review or stream data directly to a computer using the optional software and cables Each data set in...

Page 5: ...talogger option Fast response time for humidity Highly accurate microcomputer circuit Direct upload of data to Excel Detachable probe for measuring flexibility Internal clock and calendar Touch tone Tripod mounting screw Built in tabletop stand Maximum and minimum Hold function Auto power off Low battery indicator Backlight ...

Page 6: ...e powered by six AA 1 5 V UM3 batteries or an optional 9 Volt DC adapter See page 21 for battery replacement instructions Plug the adaptor into the power port labeled DC 9V located on the side of the meter Note When using the adapter to run the meter the meter will remain permanently on and the POWER button will be disabled ...

Page 7: ...7 METER COMPONENTS Top of Meter Bottom of Meter Side of Meter Type K J Thermometer Socket Humidity Temperature Probe Input Socket SD Card Socket RESET DC 9V Adapter Input Socket ...

Page 8: ...8 METER COMPONENTS Back of Meter Probe Tripod Mounting Screw Stand Battery Cover Screws Battery Compartment Cover Probe Plug Probe Head Probe Handle ...

Page 9: ...9 KEYPAD LCD Display HOLD NEXT REC ENTER LOGGER SAMPLING TIME CHECK POWER ESC Backlight SET TIME CHECK ...

Page 10: ...gger function Once selections are saved the meter will default to the selected preferences when the meter is turned off and on again Once a selection is saved the meter automatically advances to the next setup function i e after setting the real time clock the meter will automatically enter the decimal type setup function 1 Press POWER to turn the meter on 2 Press SET for 2 seconds or longer to en...

Page 11: ...clock The internal clock will function when the meter is turned off but only when the batteries have adequate power not with low battery power Decimal Type Although the decimal is commonly expressed as the symbol such as 20 6 or 1000 53 some European countries use a symbol to represent the decimal such as 20 6 or 1000 53 The meter defaults to the period symbol To adjust 1 Enter the decimal type fu...

Page 12: ...h tone function from Setup Mode see page 10 bEEP appears on the LCD 2 Press or to select yes touch tone enabled or no touch tone disabled Press ENTER to save the selection Thermometer Type K J 1 Enter the thermometer type function from Setup Mode see page 12 tYPE appears on the LCD 2 Press or to select K or J Press ENTER to save the selection Temperature Units 1 Enter the temperature units functio...

Page 13: ...at Enabling this function will format the SD memory card to work specifically with your meter Formatting the SD card will erase any previous memory on the card New SD cards should be formatted to work with your meter 1 Enter the SD memory card format function from Setup Mode see page 10 Sd F appears on the LCD 2 Press or to select yes format the SD memory card or no do not format the SD memory car...

Page 14: ...w point dP appears on the LCD Wet bulb Et appears on the LCD Type K J thermometer tP appears on the LCD Note The meter will default to the last function setting used when it is turned off and on again Humidity and Temperature 1 Press POWER to turn the meter on 2 Plug the probe plug into the probe input socket 3 Press and hold FUNC to cycle through the options until rH appears on the LCD Release th...

Page 15: ...s on the lower display Wet Bulb 1 Follow Steps 1 2 in the preceding Humidity and Temperature section 2 Press and hold FUNC to cycle through the options until Et appears on the LCD Release the FUNC button The wet bulb value C or F appears on the upper display and Et appears on the lower display Type K J Thermometer 1 Press POWER to turn the meter on 2 Press and hold FUNC to cycle through the option...

Page 16: ... Maximum and Minimum To record maximum and minimum readings 1 When measuring any parameter press REC to begin recording the maximum and minimum values REC appears on the LCD 2 Press REC The maximum value and REC MAX will appear on the LCD 3 Press REC The minimum value and REC MIN will appear on the LCD 4 To delete the maximum or minimum value press HOLD REC appears on the LCD and the meter will be...

Page 17: ... off 3 Press to turn the backlight on View Real Time Clock To view the time function during normal measurement not during datalogging 1 Press TIME CHECK The time information Year Month Date Hour Minute will appear on the lower display of the LCD View Sampling Time To view the sampling time function during normal measurement not during datalogging 1 Press SAMPLING CHECK The sampling time in seconds...

Page 18: ...time to 1 second Refer to page 13 2 Press REC REC will appear on the LCD 3 Press LOGGER REC will flash on the LCD and the alarm will sound while the measurement data and time information are saved to memory 4 To pause datalogging press LOGGER The meter will temporarily stop recording and REC will stop flashing on the LCD Press LOGGER again to resume datalogging REC will flash on the LCD 5 To finis...

Page 19: ...sition number from 1 to 99 To indicate the position location P x x 1 to 99 will appear on the lower display After selecting the position number press ENTER to confirm 4 To finish datalogging press REC for longer than 2 seconds REC will disappear from the LCD to indicate that datalogging has ended SD Card Data Structure 1 The first time a SD card is used in this meter a folder HTA01 will be generat...

Page 20: ...old 99 files A new route i e HTA02 will be generated when exceeding 99 files 4 The file s route structure HTA01 HTA01001 XLS HTA01002 XLS HTA01099 XLS HTA02 HTA02001 XLS HTA02002 XLS HTA02099 XLS HTAXX Note XX Maximum value is 10 ...

Page 21: ...ove the old batteries and replace with six new AA batteries ensuring correct polarity 4 Replace the battery cover Tighten the screws on the battery cover to secure to the meter TROUBLESHOOTING System Reset If the meter is not functioning properly i e the system is frozen and the keypad is non operational reset the meter 1 Press POWER to turn the meter on 2 Use a small tool such as a disassembled p...

Page 22: ...e K Penetration Thermometer Probe 800065 Insertion Type K Penetration Thermometer Probe 800066 Large Type K Penetration Thermometer Probe 800086 Large Type J Penetration Thermometer Probe 800073 Type K Surface Thermometer Probe 800070 Type K Surface Thermometer Probe 800072 Type K Magnetic Surface Thermometer Probe 800076 Type K Ambient Thermometer Probe 800077 Type K Beaded Wire Probe 800097 Type...

Page 23: ... or F Wet bulb Humidity C or F Type K J Thermometer C or F Temperature Compensation Automatic temperature compensation for the humidity function and the type K J thermometer Sampling Time of Display Approximately 1 second Memory Card SD card 1 GB to 16 GB Operating Temperature 0 to 50 C Operating Humidity 85 RH Power Supply Alkaline or heavy duty DC 1 5 V battery UM3 AA x6 pieces DC 9V adapter inp...

Page 24: ...is off Approximatley DC 28 mA If the backlight is on the power consumption will increase by approximately 12 mA Dimensions 7 x 2 x 2 177 x 68 x 45 mm Weight 489 g 1 lbs Humidity Range 5 to 95 RH Resolution 0 1 RH Accuracy 70 RH 3 reading 1 RH 70 RH 3 RH Temperature Range 0 C to 50 C 32 F to 122 F Resolution 0 1 Accuracy 0 8 C 1 5 F ...

Page 25: ...measurement The dew point accuracy is the sum accuracy value of the humidity and temperature measurement Wet Bulb Humidity C Range 21 6 C to 50 0 C Resolution 0 1 C F Range 6 9 F to 122 0 F Resolution 0 1 F Note The wet bulb value is automatically calculated from the humidity temperature measurement The wet bulb accuracy is the sum accuracy value of the humidity and temperature measurement ...

Page 26: ...o 1300 C 50 1 to 100 C 0 1 C 0 4 reading 0 5 C 0 4 reading 1 C 58 0 to 2372 F 58 1 to 148 F 0 1 F 0 4 reading 1 F 0 4 reading 1 8 F Type J 50 0 to 1200 C 50 1 to 100 C 0 1 C 0 4 reading 0 5 C 0 4 reading 1 C 58 0 to 2192 F 58 1 to 148 F 0 1 F 0 4 r eading 1 F 0 4 reading 1 8 F ...

Page 27: ...27 ...

Page 28: ...es not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal and void the warranty To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 8281 E Evans Rd Suite 103 Scottsdale AZ 85260 480 948 4448 The defective unit must be accompanied by a...