Operating Procedures 

1. With the COVER PLATE open, clean the PRISM 

with a soft cloth to avoid scratching the surfaces. 

2. Aim the refractometer toward a light source and 

rotate the EYEPIECE to obtain the clearest focus. 

3. Adjustment of the null (reference point): 

A. Open the COVER PLATE. 
B. Apply a few drops of pure distilled water onto 

the PRISM platform. 

C. Close the COVER PLATE. 
D. Turn the CALIBRATION SCREW until the 

dark and light boundary line coincides with the 
bottom line of the scale (calibration line). 

4.   Carefully dry the prism platform and the cover. 
5. Place a few drops of the test solution on the prism 

and close the COVER PLATE so the solution 
spreads evenly on the prism. 

6. Aim the front of the refractometer towards the light 

source and focus the eyepiece on the boundary 
line of the light and dark hemispheres. 

7. The boundary line indicates the concentration of 

Urine Specific Gravity and Refractive Index of the 
test sample. 

 

   
 
 
 
 
 
 
 

 
 
 
 

8. After use, clean prisms with a cloth and remove 

any residue. 

9. The temperature of the null reference liquid should 

be at the same temperature as the sample solu-
tion.  For variations in temperature the null point 
should be adjusted once every 30 minutes. 

 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 

—– 