Industrial Imaging System
Page 19 of 44
SPX Series System Manual
118173-001 Rev. D
System Interconnections - Air Cooled
The following details the connections for setting
up an air cooled SPX X-ray System.
! WARNING !
All cables MUST be connected
to their appropriate connectors
on the Control Unit, Cooling
Unit, and Tube Head before
applying power to the system.
1. Install the Cooling Fan power cable:
Connect the male end of the power cable
to the connector labeled
“Cooler” on the
Control Unit.
Connect the female end of the power
cable to the connector on the Tube Head
base plate.
2. Install the Tube Head Control cable:
Connect the male end of the Control
cable to the connector labeled “Tube
H
ead” on the Control Unit.
Connect the female end of the Control
cable to the connector on the Tube Head
base plate.
3. Install the line power cable:
Connect the female end of the line power
cable to the connector labeled “Power” on
the Control Unit.
Connect the “plug” end of the line power
cable to the AC voltage source (see
Connecting to Power instructions later in
this section).
4. Make the External Interlock connection:
If available, connect the Interlock cable to
the connector labeled “Interlock” on the
Control Unit (see External Interlock
instructions later in this section).
Connect the “jumper” (supplied) to the
connector labeled “Interlock” on the
Control Unit for units that
do not employ
an external interlock system.
Figure 3-2
SPX-System Setup (
air-cooled
)