SYSTEM SET UP PROCEDURES
The procedures below describe the set up
procedures for both SPX Unit configurations;
liquid-cooled units, and air-cooled units. Use
Figure 3-1 (liquid-cooled), or Figure 3-2 (air-
cooled) as a reference while making the
necessary system connections.
System Interconnections - Liquid
Cooled
The following details the connections for setting
up a liquid cooled SPX X-ray System.
! WARNING !
All cables MUST be connected
to their appropriate connectors
on the Control Unit, Cooling
Unit, and Tube Head before
applying power to the system.
1. Connect the twin hose assembly between the
Tube Head and the Cooling Unit:
_ Attach the two angled couplings to the fittings
on the back of the Tube Head.
_ Attach the couplings on the opposite end to
the female fittings on the Cooling Unit.
NOTE:
There is no designated left or right side to the
twin hose assembly. If each coupling is
properly seated, coolant flow through the
Tube Head will be achieved.
2. Install the Cooling Unit power cable:
Connect the male end of the power cable to the
connector labeled “Cooler” on the Control Unit.
Connect the female end of the power cable to the
connector on the Cooling Unit.
3. Install the Tube Head Control cable:
Connect the male end of the Control cable to the
c
onnector labeled “Tube Head” on the Control
Unit.
Connect the female end of the Control cable to
the
connector on the Tube Head base plate.
4. Install the line power cable:
Connect the female end of the line power cable to
the connector labeled “Power” on the Control Unit.
Connect the “plug” end of the line power cable to
the AC voltage source (see Connecting to Power
instructions later in this section).
5. Make the External Interlock connection:
If available, connect the Interlock cable to the
connect
or labeled “Interlock” on the Control Unit
(see External Interlock instructions later in this
section).
Connect the “jumper” (supplied) to the connector
labeled “Interlock” on the Control Unit for units
that do not employ an external interlock system.
Figure 3-1
SPX-System Setup (Liquid Cooled)
Summary of Contents for SPX Series
Page 1: ......
Page 32: ...SPX X ray Industrial Imaging System Page 30 of 44 SPX Series System Manual 118173 001 Rev D ...
Page 33: ...SPX X ray Industrial Imaging System Page 31 of 44 SPX Series System Manual 118173 001 Rev D ...
Page 43: ...SPX X ray Industrial Imaging System Page 41 of 44 SPX Series System Manual 118173 001 Rev D ...