TANGENTIAL FLOW DEPTH FILTRATION

®

(TFDF

®

)    

p 1

Product Instructions for Use

Tangential Flow Depth Filtration

®

(TFDF

®

)

Introduction

SpectrumLabs.com offers a unique TFDF

®

filter that combines

the  benefits  of  the  non-fouling  nature  of  tangential  flow  filters

with the benefits of high dirt capacity depth filters.  Please refer

to the following instructions for the care and use.

A. Materials of Construction

Housing:

Polysulfone

Filter Media:

HDPE 1-3um Pore Rating

HDPE 2-5um Pore Rating

Potting Material:

Urethane 

-

All materials USP Class VI compatible

-Pore  Ratings  for  reference  only:  Retention  rating  may  vary

based on application

B.  Filter Preparation

TFDF

®

filters  are  shipped  ready-to-use  with  no  surfactant  or

storage agent.  Filters wet readily with water and/or buffer.

C.  Filter Sterilization

SpectrumLabs.com  recommends  using  the  filters  with  Pro-

Connex

®

Ready-to-Use, Sterile Flow Paths that include the sin-

gle-use  MagLev  pump  and  pressure  sensors  for  perfusion

applications.  When needed, Sampler and MiniKros

®

filters (fil-

ters part numbers beginning with S and N) filters can be auto-

claved one time using the conditions below:  

NOTE:

KrosFlo

®

and KrosFlo

®

Max filters (filters beginning

with K or X) have not been validated for autoclave

cycles.

1.     Remove and discard all pouches, as these are not auto-

clavable; wet the membrane (optional). 

2.     Loosen all module connections to avoid damaging the

module during the autoclave cycle. 

3.     Wrap all open ports of the connecting tubing with auto-

clave paper, making sure permeate cavity is open to

steam penetration along with the inlet/retentate side. 

4.     Ramp up the autoclave by increasing the temperature by

1-2°C/min up to 121°C (15 psig)