When the meter is powered off, the calibration con-
stants are retained. Therefore, the circled indicators for
all completed calibrations will remain illuminated. How-
ever, when a new calibration is done, only the circled
indicator for that pH level will be displayed. If a multiple-
point calibration is desired, the calibration must be per-
formed for the other pH levels as well.
Notes:
1. If the batteries are removed or go dead, any user-
created calibration data will be lost and the meter
will return to the factory defaults.
2. Always turn the meter off and then on before cali-
brating to allow sufficient time to complete the cali-
brations during one power cycle, especially for
multi-point calibrations. If the meter powers off
automatically during a calibration, any completed
calibrations will be saved. But, after the meter is
powered up again, any new multi-point calibration
will have to be done in entirety.
Changing the
Temperature Units
Press and hold the
CAL/RECALL
button for approxi-
mately 3 seconds. The
o
C or
o
F symbol will change
first and the numerical temperature value will change
after
the button is released. If the meter accidentally
transitions into
CAL
mode, the
CAL
symbol will appear
on the LCD. Simply turn the SoilStik off and start again.