13
Display Removal:
1.
Follow steps for
Display and Sensor Removal
to access the
display and cable connections.
2.
Remove the foam spacer and unplug the sensor cable connect-
or from the jack. If an IR temperature sensor is connected,
disconnect this plug as well.
Do not discard the foam spac-
er.
Display Installation:
1.
Follow the procedures for
Display Removal
to access the dis-
play and cable connections.
2.
Re
-
connect the sensor cable to the 3.5mm connector on the
back of the display module.
3.
Insert the foam spacer back into place behind the sensor cable.
The forked end fits around the sensor cable molding.
4.
If equipped with the IR temperature sensor option, attach it to
the smaller diameter connector.
5.
Guide the excess cable length back down through the base
plate.
6.
Mount the display back onto the base plate by tightening the
four screws. Be sure the USB port side faces the same side as
the serial number label.
Sensor Block Removal / Replacement:
1.
Follow steps for
Display Removal
to access the cable con-
nections.
2.
Remove the fastening bolt that joins the probe block to the
shaft.
3.
Separate the probe block from the shaft (fig. 2).
4.
Feed the cable from the replacement probe block through the
shaft. A string fed down from the top can be used to aide in
the process of cable reconnection.
5.
Follow the procedure for Display Installation to complete the
replacement.