14
Sensor/Tip Replacement
If the probe cannot be calibrated, or if it does not
hold the calibration for a reasonable amount of time,
the probe or tip must be replaced. The
24”
T
-
handle
probe comes with a replaceable tip (item 2266). With
these probes, the cable and shaft do not need to be
replaced.
An
“
Temperature out of Range
”
message on the LCD
may indicate a failed probe or simply that the probe
has come unplugged from the meter. See
Attaching
a direct
-
insert probe
(p. 15) before replacing the
probe or tip.
Replacing a tip
(24”
probes only)
The replaceable tip (left) is designed to thread
in and out of the end of the probe. A new tip
has two flat faces on opposite sides. This al-
lows the tip to be tightened onto the probe with
a pliers or
5/16”
wrench.
I. Removing a direct
-
insert probe
The
8”
probe do not have a replaceable tip. The re-
placement procedure is outlined below.
1.
Remove the black nut from the strain
-
relief
where the cable feeds into the meter.
2.
Remove the white nut on the cable end of the
meter.
3.
Detach the probe cable plug from the socket in
the meter.
4.
Remove the white nut and large o
-
ring.