6. Press HOLD to accept the auto conductivity
standard and finish the calibration.
7. Display returns to Measurement window.
Manual Calibration
When the conductivity reading is outside the
calibration range of the automatic conductivity
standards or when TDS or salinity is used, the tester
will require manual adjustment.
1. Repeat steps 1 to 4 from
“
Automatic Calibration
”.
2. Wait until the display settles on a measured value.
Press MENU/
▼
to manually adjust the value to the
desired reading.
Note:
The adjustment will decrease only, however
the adjustment will eventually cycle to the highest
available value after decreasing by 40% of the initial
value.
3. Press HOLD
to accept and finish the calibration
when the desired value is selected.
Note:
To abort calibration, press CAL/ESC to
escape.
4. Once the calibration is finished and user has
accepted the changes, the Measurement window will
show the calibrated reading.