Spectros Medical Devices Inc.
Page 34 of 44
Rev B
T-STAT 2.0 User Manual
T-Stat® Sensor Specifications applicable to all sensors
1. Physical Specifications
Length:
3 m Long
Weight:
2 oz (60 g)
2. Optical Specifications:
Connector Output:
Fiber optic return line
Connector Input:
Internally-current-limited electrically-isolated 5v electrical source
Connector:
Male, Type 1 Spectros Light-Jack Connector
Light Output:
Non-laser broadband LED source
Visible wavelength range (420-700) nm
Luminous Intensity 12.9 cd TYP
3. Electrical/Emissions Specifications
:
Input Voltage:
5V DC
Electrically isolated
Power Required:
500 mVA (typical)
Battery:
Back-Up
Fuse:
None
Patient/Operator Protection:
Isolated 5V DC supply to sensor socket
Dual current limit resistor (socket and sensor)
Memory
Internal EEPROM 1kB memory, with the following information:
–
Lot Number
–
Serial Number
–
Mfg Date
–
Sensor Name
–
Sensor Type
–
Compatibility Version
–
Expiration Date
–
Expiration Flag
–
Reuse Flags
–
Concentration and/or wavelength calibration factor
Approvals:
Tested with T-Stat 2.0 to the following specifications:
UL 2601-1 ; EN60601-1(2005) Class 1, Type BF, continuous operation ; IEC 60601-1-2
(2007),
4. Operating Specifications
:
Environment:
Intended for Indoor Hospital Use
5-35 degrees Celsius
5%-95% humidity (non-condensing)
Start-Up
Warm-up and self test requires up to 2 minutes
5. Sensor Specifications
:
Sterile
Sensors supplied sterile for single use.
6. Storage Specifications
:
Temperature/humidity range
–
not
to exceed
-20-60 degrees C 5-95% RH
Storage
We recommend WHO recommendation of 15-25C Ambient. Temperature can vary up to the -
20 to +60C limits so long as long-term average is 25C or less.