![Spectris Particle Measuring Systems Airnet II 201-4 Operation Manual Download Page 24](http://html.mh-extra.com/html/spectris/particle-measuring-systems-airnet-ii-201-4/particle-measuring-systems-airnet-ii-201-4_operation-manual_1344041024.webp)
Installation
Chapter 2
: Unpacking and Installation
Airnet® II 4-Channel Particle Sensors Operations Manual
2-5
Powering the Sensor from 24 VDC
The Airnet II 4-Channel particle sensor can be powered with sources other than the
Power over Ethernet (PoE). The unit may also be powered using a source of 24 VDC
power (if this option was ordered) that may be connected to multiple Airnet II 4-
Channel particle sensors. Ensure that the 24 VDC supply wiring is sized 18 AWG (0.75
mm
2
) minimum and the supply is current limited to 8 A maximum or fused at 5A
maximum.
If using a local power source a receptacle and crimp pins are available that attach to
the ends of the power cable. The crimp pins are then inserted into the provided 12-pin
Molex™ Micro Fit receptacle.
Figure 2-5
24 VDC Power Connector
Powering the Sensor from Power over Ethernet (PoE)
The Airnet II 4-Channel can be powered from a Power over Ethernet Router or Injector
that conforms to IEEE 802.3af. A shielded CAT5 cable must be used or a local ground
connected to the sensor for proper operation when using PoE.