2-2
Airnet® II 4-Channel Particle Sensors Operations Manual
Chapter 2
: Unpacking and Installation
Items You Need to Provide
Items You Need to Provide
The following items are not included, and you will need to provide:
•
Ethernet cable (shielded CAT 5 UTP cable if PoE is used)
•
4-20 mA cable (necessary only if you will use 4-20 mA signal output)
•
Computer
•
Facility Net software
•
Modbus TCP client software
•
Power cables, IEC type C 13, necessary for local power
Installation
The following instructions are included to guide you through the installation process.
Please pay close attention to the
NOTE
,
CAUTION
, and
WARNING
text.
Installation includes the following steps:
•
Selecting an Appropriate Location
•
Positioning or Mounting the Unit
•
Connecting the Power Supply
•
Connecting a Communications Cable
•
Connecting Sample Tubing or Sample Probe (as applicable)
•
Attaching Vacuum Tubing
Automated Laser Power Control
Airnet II 4-Channel particle sensor units contain an automated laser power control that
automatically shuts off the laser power after 10 seconds of continuous bad flow (flow
rate that is ± > 15% of the specified flow rate).
The laser power is restored after 10 seconds of continuous good flow.
Detector Board Monitor
The Airnet II 4-Channel particle sensor has an automatic monitor to assure the
detector board is operating correctly. In many applications the normal condition is
no
particle counts being detected
for an extended period of time.
NOTE:
If installation of the Airnet II 4-Channel particle sensor will
prevent access to its RS-232 port (needed for configuration),
configure the particle sensor first and then install the unit. See