System Setup
TM 799200 Rev. (4), April 2017
35
3.3.2.6
Heated Optic Operation
The heated optics for the detector unit can be defined as 1 of the following
modes:
Off – Not operated
On – Operated continuously
Auto – On, per temperature change (default)
When the detector is set “per temperature change,” the user can define the
start temperature below which the window will be heated, to be between 0–
50°C / 32–122°F. The detector will stop heating the window when the
temperature reaches 15°C/59°F above the defined temperature.
This feature applies only to the detector.
The source heated optic must be defined with the order, as 1 of 2 options:
Heated continuously
Or
Start heating below 41°F/5°C (default)
3.3.3
Default Detector Setup
The detector has 8 functions that can be programmed according to customer
requirements, either at the factory or at the customer facility, using a PC
software host or a handheld unit. The standard setup is as follows:
Table 9: Detector Default Setup
Function
Setting
Gas Type
2
Full scale sensitivity
5 LEL.m
BG Zero Calibration
Enabled
Accessory Relay
Warning*
Alarm Latched
No
0–20mA
Continuous**
Beam block during alarm
No latch
Heat mode
Auto
Heat on
5
*Only refers to wiring option with accessory relay output
** Only refer to the wiring options with 0–20mA
Summary of Contents for SafEye Xenon 700S
Page 2: ...This page intentionally left blank ...
Page 12: ......
Page 16: ......
Page 24: ...Models and Types 24 SafEyeTM Xenon 700S Gas Detector User Guide Figure 1 Wiring Options ...
Page 64: ......
Page 72: ......
Page 76: ......