System Information

Fig. A: System Diagram

Purity Monitor  
EPM-4

DI Cartridge #1
DI-SFCC-10

DI Cartridge #2
DI-SFCC-10

DI Cartridge #3
DI-AR-CI-10

DI Cartridge #4
DI-AR-CI-10

Monitor Probe
in Tee

Fig. B: Purity Monitor Without Cover

SpectraPure

®

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

9

Troubleshooting the Water Purity Monitor

Some Frequently Asked Questions

Q:Does monitor read “good” or “bad” when it is unplugged ?
A:The monitor reads “bad-RED” when it is unplugged.

Q: What if the light switches between GREEN and RED between 
successive depressions of the “push-to-test” water purity monitor?
A: Do not press the “push-to-test” button rapidly in quick succes-
sion. Wait at least 1 minute between two successive depressions 
of the button for an accurate reading.

Q:How should the probe be mounted for an accurate reading?
A:The probe should be mounted VERTICALLY, at the bottom of 
the T.

Q:How long does the battery last ?
A: approx 1 year

Q:How do you determine if the battery needs replacement ?
A:Depressing “push-to-test” button does not give any light.

Q: If the monitor shows a red light, what does it mean?
A: It means that either the membrane is bad OR either the probe 
or monitor is faulty.

Monitor Testing

4.  Install a new cartridge (DI-AR-CI-10) into Stage #4. Re-install the 
 

bottom housings onto the caps by rotating them clock-wise and 

 

tighten with the wrench.

5.  If no other service is required, turn the system on and check for 
 leaks.

Summary of Contents for The Ultimate DI System

Page 1: ...ro rated the performance of membrane elements for one year from date of receipt by the buyer providing that the loss of performance was not caused by fouling neglect or water conditions exceeding the feed water parameters listed on page 3 of this manual Most municipal water supplies meet these requirements Terms and Conditions of Sale 1 To obtain service under this warranty the defective system or...

Page 2: ...nates contaminant leakage from exhausted DI cartridges with the addition of secondary back up cartridges An exhausted deionization cartridge will leach the removed impurities back into the water at levels many times higher than originally found in the source water The addition of the two secondary DI stages prevents this leakage When the first SilicaFree Cartridge DI SFCC 10 changes color upon exh...

Page 3: ... contaminated product water due to exhausted cartridges is completely eliminated With the majority of contaminants now removed by Stages 1 and 2 the water is then passed through two additional stages of DI resin Stages 3 and 4 have our Chloramine Removal Car tridges DI AR CI 10 They will provide longer life than standard mixed bed cartridges when used to treat water containing chlorine chloramine ...

Page 4: ...tween GREEN and RED between successive depressions of the push to test water purity monitor A Do not press the push to test button rapidly in quick succes sion Wait at least 1 minute between two successive depressions of the button for an accurate reading Q How should the probe be mounted for an accurate reading A The probe should be mounted VERTICALLY at the bottom of the T Q How long does the ba...

Page 5: ...Needed One DI AR CI 10 Deionizaton cartridge filter wrench Procedure 1 Remove the filter housing from Stage 3 by unscrewing it counter clockwise as viewed from the bottom 2 Remove and discard the old cartridge from the housing 3 Remove the housing and cartridge from Stage 4 and re install the cartridge in Stage 3 Make sure the cartridge is installed in the correct direction as marked on the filter...

Page 6: ...rapure com Visit us on the web www spectrapure com 7 The Ultimate DI System has been engineered to provide fool proof operation and double assurance that contaminants will not be intro duced into the product water except in cases of extreme negligence The design theory to remember is that the cartridges will be replaced in a rotational pattern Stages 1 and 3 will always exhaust their capacity firs...

Reviews: