Troubleshooting the Water Purity Monitor
Some Frequently Asked Questions
Q:Does monitor read “good” or “bad” when it is unplugged ?
A:The monitor reads “bad-RED” when it is unplugged.
Q: What if the light switches between GREEN and RED between
successive depressions of the “push-to-test” water purity monitor?
A: Do not press the “push-to-test” button rapidly in quick succes-
sion. Wait at least 1 minute between two successive depressions
of the button for an accurate reading.
Q:How should the probe be mounted for an accurate reading?
A:The probe should be mounted VERTICALLY, at the bottom of
the T.
Q:How long does the battery last ?
A: approx 1 year
Q:How do you determine if the battery needs replacement ?
A:Depressing “push-to-test” button does not give any light.
Q: If the monitor shows a red light, what does it mean?
A: It means that either the membrane is bad OR either the probe
or monitor is faulty.
Monitor Testing
4. Install a new cartridge (DI-AR-CI-10) into Stage #4. Re-install the
bottom housings onto the caps by rotating them clock-wise and
tighten with the wrench.
5. If no other service is required, turn the system on and check for
leaks.