background image

SpectraPure

SpectraPure®Inc.

  

480.894.5437   Call us toll-free  1.800.685.2783    

2167 East Fifth St,    Tempe, Arizona   85281

®

10

System Hook-Up

1.  Attach the garden hose adapter to your cold water source. Never run hot    
 

water (greater than 100° F (38° C)) through the system.

2.  Place the yellow concentrate tubing and the blue purified water tubing into   
 

a drain. Do not restrict flow from these lines.

3.  Slowly open the cold water supply valve and allow the  housings to  fill. You    
 

may use pressure up to 80 psi (5.5 bar). 

Check the system to ensure that all fittings are tight and leak-free before 
leaving the system unattended.

Note: It is recommended that at least 2 gallons (7.57 liters) of purified water be 
discarded before collecting purified water for use. If the unit is not used for sev-
eral days, run the system for at least 15 minutes before collecting any water.

Initally, air trapped in the DI cartridges is a normal condition and will not affect 
the operation of the DI cartridges.

4.  Finish the installation by connecting the yellow line to a drain. The blue line
 

 can be placed into a RubberMaid™ trash can or you can set it up for 

 

automatic shut-off using a Liquid Level Controller (LLC-S-115 or LLC-M-115)    

 

or the Auto Shut-Off Kit (ASOFK).

SYSTEM INITIALIZATION

If you are setting up your system for the first time or replacing the RO 
membrane, please see the “Checking the Concentrate to Purified Water 
Ratio” section on page 11-12 .

SpectraPure®Inc

. assumes no responsibility for water damage 

due to leaks. It is the user’s responsibility to determine that the 

system is leak-free.

SpectraPure®Inc

 Fax  480.894.6109 Fax us toll-free 1.877.527.7873

E-mail: [email protected] Visit us on the web www.spectrapure.com

19

SpectraPure

®

RO MEMBRANE REPLACEMENT 

1.  Turn off the water supply to the RO system. Place the system   

 

 

where the membrane housing is easily accessible.

2.  Remove the black tubing from the membrane feed push-fitting 

 

by depressing the collar on the fitting with your thumb and pulling  

 

the tubing from the push-fitting (Page 13).

3.  Lift the membrane housing from the retention clips.
4.  Unscrew the membrane housing lid. This may require two people.
5.  Use a pair of pliers to grasp the membrane stem and pull the    

 

 

membrane from the housing (Fig. H).

6.  Remove the black housing O-ring (Fig. H). Wash the empty housing 

 

with soapy water. Rinse thoroughly with hot, clean water.

7.  Insert new membrane into the housing, with the double

 

O-ring end first (Fig. I). The tube must fit into the recess at the 

 

bottom of the membrane housing. When the membrane is aligned 

 

with the hole, firmly push the membrane into the hole until it 

 bottoms 

out. 

8.  Place the black housing O-ring on the housing rim and carefully 

 

screw the lid back on to the base.

9.  Reconnect the black tubing to the membrane feed push-fitting. 
Note: If you have a dual-membrane system, perform steps 2 thru 9 on

the second membrane now.
10.  Disconnect the yellow concentrate tubing (Fig. B) from the    

 

 

membrane housing and remove the flow restrictor (Fig. C) from the  

 

yellow tubing (Refer to the procedure on page 13-14). 

Fig. G: Membrane Housing

blue product 
water tubing

black feed tubing

membrane feed 
push-fitting

membrane housing lid

yellow concentrate tubing

Summary of Contents for SpectraPlus 2000

Page 1: ...trictor for 180 gpd 680 lpd System DI SFCI 10 SilicaFree Color Change DI Cartridge DI AR CI 10 Chloramine Removal Cartridge GHA 4 1 4 6 35 mm Garden Hose Adapter XWR UNIV Filter Wrench Optional Accessories Model Optional Part FAU SMP Quick Connect Faucet Coupler TS C61 Micro Siemens Conductivity Tester 0 1999 uS cm TS T71 Total Disolved Solids Tester 0 1999 ppm TDS TK CL 25 Total Chlorine Test Kit...

Page 2: ...18 20 Troubleshooting Guide Troubleshooting Guide for the System 21 Troubleshooting Purity Monitor 22 Troubleshooting the RO Membrane 23 26 Mounting Locations 27 Replacement Optional Parts 28 SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 27 SpectraPure CHOOSING A MOUNTING LOCATIONS When considering a location...

Page 3: ... specified 3 This warranty is in lieu of all other warranties expressed or implied including warranties of fitness for a particular purpose 4 We do not authorize any person or representative to assume for us any other oblligation on the sale of our equipment This is the exclusive remedy and liability for consequential damages under any and all warranties which are excluded to the extent exclusion ...

Page 4: ...quipment and all returned goods must also have adequate insurance coverage and a tracking number SpectraPure will not pay for loss or damage caused directly or indirectly by the presence growth proliferation spread or any activity of fungus wet or dry rot or bacteria Such loss or damage is excluded regardless of any other cause or event that contributes concurrently or in any sequence to the loss ...

Page 5: ...ost salts and other impurities are rejected and automatically rinsed from the membrane down the drain Figure 1 Figure 3 WHAT IS REVERSE OSMOSIS Osmosis is a process in nature that allows fluid of a lower concentration to pass through a semi permeable membrane into a fluid of a higher concentration See Figure 1 Because of the difference in salt concentration pure water flows through the membrane as...

Page 6: ...30 more than 49 GPD use BPHF MO 115 230 Operating pressure greater than 80 psi will require a pressure reducing valve SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 23 SpectraPure TESTING THE QUALITY OF THE MEMBRANE Membrane Output Calculation Membranes produce the rated gallons per day GPD at 60 psi 4 1 bars ...

Page 7: ...lter This filter is required to remove excess turbidity particulate matter that may cause the carbon block filter to plug 2 The second stage of filtration is a 0 5 micron carbon block pre filter This filter removes organics and chlorine from the feed water that can damage the membrane 3 The third filtration stage of the system is a high rejection thin film composite TFC reverse osmosis membrane It...

Page 8: ...STEMS 1 Low production rate a plugged pre filters i Replace pre filters b low water temperature ii Heat feed water OR use higher GPD membrane c low line pressure iii Use booster pump OR use higher GPD membrane d high TDS content iv Use booster pump OR use higher GPD membrane e fouled membrane v Clean or replace membrane to restore flux f plugged flow restrictor vi Replace flow restrictor membrane ...

Page 9: ... the web www spectrapure com 9 SpectraPure Fig C Purity Monitor Without Cover Switch Number 1 2 3 4 5 6 7 8 Setting in uS 100 50 20 10 5 2 1 0 5 Setting in ppm 67 33 13 6 7 3 3 1 3 0 67 0 33 Fig B Purity Monitor Switches If settings are changed for any reason such as testing the quality of the RO membrane They should be returned to the original 7 setting after completion of the test Switch 7 Shown...

Page 10: ... is the user s responsibility to determine that the system is leak free SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 19 SpectraPure RO MEMBRANE REPLACEMENT 1 Turn off the water supply to the RO system Place the system where the membrane housing is easily accessible 2 Remove the black tubing from the membrane...

Page 11: ...eceive the most accurate mea SpectraPure Inc Fax 480 894 6109 Fax us toll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 11 SpectraPure CHECKING THE CONCENTRATE TO PURIFIED WATER RATIO This procedure will assure you of maximum life and reliability of your SpectraPure System Failure to perform this procedure can permanently damage the membrane and will vo...

Page 12: ...MENT Right most housing The last stage is a dual layer cation anion cartridge or in some cases a mixed bed DI The condition of this cartridge should be judged by observ ing either a RED or GREEN light indication from the monitor Please note that the probe is located after housing 4 Thus the monitor indicates the condition of the final DI stage cartridge only Turn the on the system and allow water ...

Page 13: ...oll free 1 877 527 7873 E mail spectra spectrapure com Visit us on the web www spectrapure com 13 SpectraPure FLOW RESTRICTOR REMOVAL ADJUSTMENT AND REPLACEMENT 1 Locate the yellow concentrate tubing Fig E Remove the tubing from its push fitting at the membrane as follows a Firmly depress and hold the push fitting collar down with your thumbnail b While the push fitting collar is depressed pull th...

Page 14: ...he sediment and carbon pre filters must be changed when a 15 20 drop in pressure is observed OR at least 6 month intervals If your water contains a great deal of sediment or chlorine the pre filters may require more frequent changes to maintain adequate production rate and extended membrane life Sediment Pre Filter Replacement Materials Required 0 5 micron MicroTec Sediment Filter SF MT 0 5 10 Fil...

Reviews: