PSR-1100

f

 

Portable Spectroradiometer 

Operator’s Manual 

 
 

 

 

Summary of Contents for PSR-1100f

Page 1: ...PSR 1100f Portable Spectroradiometer Operator s Manual...

Page 2: ...USB or Bluetooth Virtual COM Port 7 2 2 4 ESD protection 7 3 Control Panel 8 3 1 Indicators 8 3 2 Control Panel Switches 8 3 2 1 Power Switch 8 3 2 2 Membrane Switches 8 3 2 3 Switch Function 8 3 3 LC...

Page 3: ...1100f system operation are built into the instrument The instrument incorporates manual controls and on board storage for self contained field operation as well as the capability to connect to a host...

Page 4: ...ut FOV The entire spectrometer foreoptic system is factory calibrated for radiance using a NIST traceable source The calibration coefficients are stored in flash memory by the internal microprocessor...

Page 5: ...100f contains the on off switch USB connector battery receptacle and a control panel with membrane switches and LCD display On either side of the front panel there are two mounting brackets for attach...

Page 6: ...iew of PSR 1100f optical components 2 2 Interface Specification 2 2 1 Battery Pack Your PSR 1100f is supplied with one or more Li Ion battery packs that fit into the recess on the front panel Slide th...

Page 7: ...l Connector For communications via USB the front panel is equipped with a USB Type B receptacle connector compatible with all standard USB cables with Type B plug Pin 1 5V Pin 2 D Pin 3 D Pin 4 GND 2...

Page 8: ...y of modifying the unit s operation The functions of the switches are detailed below 3 2 3 Switch Function Scan Steps through scan options Reference Target Dark Scans Menu Steps through the menu optio...

Page 9: ...l hold the next acquired scan The allowable values are from 1 to 500 The memory slot cannot be adjusted by the keypad with the exception of completely erasing all memory To erase all memory and set ME...

Page 10: ...28 17 3 3 4 Modifying a Parameter 1 Depress the MENU button to locate the parameter you wish to edit 2 Modify the parameter by depressing the EDIT button The push button switches are programmed to sc...

Page 11: ...en manually by the operator when in Scaled Dark Modes In AutoDark Mode the internal shutter automatically closes and dark measurement taken immediately prior to the reference or target scan there is n...

Page 12: ...s established Dark measurements are stored in volatile memory not flash and must be retaken whenever the instrument power is cycled Taking a REFERENCE Measurement 1 Check that the LED Power Indicator...

Page 13: ...in real time with expanded control options and Retrieving and displaying scan data collected and stored during standalone operation Be sure to install the software and any USB Bluetooth communications...

Page 14: ...inated Charging a battery from empty to full capacity approximately 8 4VDC will take approximately 2 3 hours 5 3 Information and Warnings Cells 18650 Li ION cells with IC protection are used inside th...

Page 15: ...cy 5 nm Wavelength Reproducibility 1 nm Integration Time 8 2000 ms Physical Specifications Power Supplied 2800 mA h Li Ion battery pack with charger 7 5V nominal Size 7 x 3 25 x 5 75 Weight 4 lbs Comm...

Page 16: ...reference material for reflectance scans The plate is made of polytetrafluoroethlyene PTFE and has high diffuse reflectance Over time the surface of the plate may become dirty or scratched In order to...

Page 17: ...te 1 00 Original Draft 4 5 11 1 01 Added Appendix D on White Reference 5 16 12 1 02 Updated specifications 5 29 12 1 03 Updated mailing address 11 15 12 1 04 Added CE Compliance Declaration PSR 1100F...