Acquisition Control
After priming of the cartridge is complete, continuous data acquisition will begin automatically
using bias values that are optimized for the cartridge type.
Continuous-mode and Single-mode Data Acquisition
➢
The length of each acquisition is specified using the “Acquisition Time” dropdown (10s
default).
➢
Continuous-
mode acquisition may be manually started by pressing the “Continuous”
button (Figure 7).
Repeat acquisitions are collected until the user presses the “Stop”
button
or until user-
defined “Set Acquisition Stop Point” value is reached, see page
➢
In continuous mode, each acquisition is automatically flagged for processing unless a
significant number of defects are detected in the data (see page 18).
➢
A sin
gle acquisition can be obtained by pressing the “Single” button.
➢
In single acquisition mode, the user may manually flag the most recent acquisition for
automatic processing by pressing the “ON” button in the Acquisition frame.
➢
Interrupt the current acqui
sition with the “Stop” button. Partial acquisitions are not saved.
Figure 7.
Acquisition frame.
Data Saving Tips:
➢
Raw data files are saved to the folder specified by the user in the Run
Details frame.
➢
Raw data file names have the following format:
❖
<prefix>_YYYYMMDD_NNNN.h5
➢
Record significant acquisition numbers in your notes for future
reference.