Note:
If the Silver Gate™ bridge and clamp anvil assembly (
15
) is
removed by loosening and removing the hinge pin, take care
that the stay up catch support rod (
22
) that is loose fitted in the
bridge and which engages against the stay up catch device (
19
)
does not fall out of the bridge and becomes lost (See Figure 19.)
Figure 19. Silver Gate™ Accessory with Bridge (15) up showing
detail of the stay up catch (19) and stay up support rod (22)
For analyzing a liquid or a paste:
1. Collect a background spectrum as described in steps 1 to 4 (page
21).
2. If the Silver Gate™ ATR Accessory is fitted with a bridge and clamp
anvil assembly (15), open the bridge and position it past the stay up
catch device (
19
). This allows for easier access to the ATR crystal
surface area to position the sample correctly.
3. Take the liquid or paste sample and place it very carefully such that
it covers the entire surface of the ZnSe, Ge or Si ATR crystal. A
19
22
15