Monitor Report
Condition
EC09
EC19
Contradictory instructions sent to hardware (e.g., “pump ON and valve
open”).
EC29
Diastolic pressure value cannot be obtained from the data available.
EC39
Algorithm could not process input data quickly enough, resulting in an input
queue overflow.
EC49
This monitor must be initialized.
EC59
Heart rate value cannot be obtained from the data available.
EC69
Heart rate value cannot be obtained from the data available.
EC79
Bleed steps were too small. This may be caused by a partially obstructed
air hose. All blood pressure attempts following this message are inhibited.
Attempts can be enabled by powering the monitor OFF and then ON.
EC99
Unexpected or contradictory data (such as a negative cuff pressure).
LLL
EC16
Low battery detected prior to start of measurement.
EC26
Low battery detected after measurement started. Usually caused by the
pump drawing enough current to lower the battery voltage.
Lbb
The report does not print an event code for this condition, which is a low
backup battery. Contact Spacelabs Healthcare for replacement of the
battery.
Summary of Contents for 90217A
Page 4: ...ABP 90217A Operations Manual iv This page is intentionally blank...
Page 24: ...ABP 90217A Operations Manual 1 18 This page is intentionally blank...
Page 36: ...ABP 90217A Operations Manual 3 10 This page is intentionally blank...
Page 44: ...ABP 90217A Operations Manual A 6 This page is intentionally blank...
Page 45: ......