Monitor Report
Condition
EC15
Equipment malfunction. Return to Spacelabs Healthcare for service.
EC25
Monitor failed to initialize. Please initialize.
EC35
The monitor needs to be reinitialized.
EC05
EC05 &
EC45
EC45
Invalid bleed size. The monitor has automatically changed the bleed size to
8mmHg.
EC05 &
EC55
EC55
An unexpected loss of power possibly caused by: a) removal of the batteries
during a blood pressure measurement, b) hardware overpressure, or c) a
hardware time-out. Frequent EC55 messages would indicate a malfunction
which requires service.
EC05 &
EC65
EC65
Extremely large artifact.
EC05 &
EC75
EC75
Equipment malfunction. Return to Spacelabs Healthcare for service.
EC05 &
EC85
EC85
Equipment malfunction. If EC85 occurs replace batteries and use as normal
if problem persists return to Spacelabs Healthcare
EC05 &
EC95
EC95
Cuff pressure baseline is out of limits. The monitor should correct the
baseline automatically within 10 minutes, or it can be reset by initializing
the monitor. If initialization does not correct the condition, the monitor
must be returned to Spacelabs Healthcare for calibration.
EC07
EC78
Clogged connector filter.
EC08
EC18
Too few data entries to accurately determine blood pressure. This message
may indicate that the cuff is not being worn by the patient (taken off but
left connected to the monitor). The message may also indicate that motion
artifacts cause the majority of the incomplete data.
EC28
Diastole above 200mmHg.
EC38
Pulse pressure less than 16mmHg.
EC48
A) Movement artifact at mean arterial pressure
B) Heart rate arrhythmia
EC58
A) Movement artifact at diastole
B) Heart rate arrhythmia
EC68
Division by zero.
Summary of Contents for 90217A
Page 4: ...ABP 90217A Operations Manual iv This page is intentionally blank...
Page 24: ...ABP 90217A Operations Manual 1 18 This page is intentionally blank...
Page 36: ...ABP 90217A Operations Manual 3 10 This page is intentionally blank...
Page 44: ...ABP 90217A Operations Manual A 6 This page is intentionally blank...
Page 45: ......