Door Gasket
To inspect the door gasket:
Remove the door gasket.
1.
Grasp the gasket in both hands and turn it inside out.
2.
Inspect the inside of the gasket. This is where you are most likely to find a
3.
potential problem.
If any inconsistencies are found (
e.g.
, cuts, cracks, dry rot, rippling) on the
4.
inside of the gasket, it should be replaced immediately.
Note:
For information regarding a replacement door gasket, contact SP Scientific
.
Always install door gaskets on a clean, grease-free metal rim. Remove
5.
excess grease from the metal using isopropyl alcohol on a fresh paper towel
or cloth. The gasket may also be cleaned with isopropyl alcohol.
Once the new or cleaned gasket is installed, apply a very thin coating of high
6.
vacuum grease to the outer surface of the gasket only. The gasket should
appear slightly moist.
Summary of Contents for VIRTIS ADVANTAGE PRO
Page 1: ...FREEZE DRYER OPERATOR S MANUAL VIRTIS ADVANTAGE PRO LABORATORY BENCHTOP FREEZE DRYER...
Page 2: ......
Page 6: ...iv Rev 002 01 14 SP Scientific 2014...
Page 10: ......
Page 14: ...Introduction VirTis AdVantage Pro 4 Rev 002 01 14 SP Scientific 2014...
Page 32: ...Additional Options VirTis AdVantage Pro 22 Rev 002 01 14 SP Scientific 2014...
Page 46: ...Appendix B Cover Removal VirTis AdVantage Pro 36 Rev 002 01 14 SP Scientific 2014...
Page 50: ...VirTis AdVantage Pro 40 Rev 002 01 14 SP Scientific 2014 Appendix D Serum Vial Capacities...
Page 52: ...Appendix E Stainless Steel Cleaners VirTis AdVantage Pro 42 Rev 002 01 14 SP Scientific 2014...
Page 77: ...VirTis AdVantage Pro Appendix F Chemical Resistance Charts Rev 002 01 14 67 SP Scientific 2014...