22
Southworth Products
Mortuary Lift Manual
9.3 Battery Charging
Risk of explosive gases. Always charge in a well ventilated area. Keep
open flames, sparks, and metal objects away from the battery at all
times.
Do not use with an extension cord or adapter.
Always wear proper personal protective equipment when working on or
around the battery.
This charger must only be used to charge lead acid or AGM type
batteries. Use with other battery types may cause the battery to explode
and cause severe injury or death.
Always unplug the charger from the outlet before performing any
maintenance on the battery.
This charger must not be exposed to dirty or wet conditions.
Do not operate the lift while the charger is plugged in.
Charger Description Automatic 3-Stage On-Board
industrial battery charger
AC Input:
115 V, 1 Ph, 60 Hz @ 1.6 Amps
DC Output:
12 VDC @ 6 Amps
Led Indicators
On
Off
Condition
Red
Green
Charger is in the “Soft Start” or “Bulk Stage” and the battery
is being charged.
Red &
Green
None
Charger is in the “Absorption Stage” and delivering constant
voltage to the battery.
Green
Red
The charger has moved to the “Float Stage” and is topping
off the charge to the battery and keeping the battery ready to
use. The green light indicates your battery is ready to use.
STATUS LEDS
3/8” RING
TERMINALS
IN LINE FUSE HOLDER
FUSE: BUSSMAN AGX-10
(OR EQUIVALENT)
115 VAC POWER CORD