Potential danger. Indicates the user must refer to the

manual for important safety information
Indicates hazardous voltages may be present

Equipment is protected by double or reinforced insulation

Indicates the terminal(s) so marked must not be connected to a circuit where the

voltage with respect to earth ground exceeds the maximum safety rating of the meter 

Indicates the terminal(s) so marked may be subjected to hazardous voltages.

MAX

1000V

International Safety Symbols

Brief Description

Typical Applications

Category

Rating

Single phase receptacles 

and connected loads 

Three phase circuits and 

single phase lighting 

circuits in commercial 

buildings 

- Household appliances, power tools

- Outlets more than 30ft (10m) from a CAT III source

- Outlets more than 60ft (20m) from a CAT IV source
- Equipment in fixed installations such as 3-phase 

  motors, switchgear and distribution panels 

- Lighting circuits in commercial buildings

- Feeder lines in industrial plants

- Any device or branch circuit that is close to a CAT III source

CAT II

CAT III

The measurement category (CAT) rating and voltage rating is determined by a combination of the meter, test probes and any

accessories connected to the meter and test probes. The combination rating is the LOWEST of any individual component.

CAT IV

Connection point to

utility power and

outdoor conductors 

- Primary distribution panels

- Overhead or underground lines to detached buildings

- Incoming service entrance from utility

- Outdoor pumps

WARNING: 

                                   Operation is limited to CAT II applications when the insulated tips are 

removed

from one or both test probes. Refer to Input Limits section of this manual for maximum voltage ratings.

CAT III 1000V

CAT IV 600V  

CAT II

1000V

Insulated Tip On

Insulated Tip Removed

4

Maintenance

This Multimeter is designed to provide years of dependable service,

if the following care instructions are performed:
1. KEEP THE METER DRY. If it gets wet, wipe it off.
2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature

    extremes can shorten the life of the electronic parts and distort or melt

    plastic parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the

    electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth.

    DO NOT use chemicals, cleaning solvents, or detergents.
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE.

    Remove old or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the batteries

    should be removed to prevent damage to the unit.

Meter Description

1.

5.

3.
4.

10.

11.
12.

8.

9.

6.

7.

1.  LCD display 
2.  MODE select button
3.  Rotary function switch
4.  10A (positive) input jack
5.  µA/mA (positive) input jack
6.  COM (negative) input jack  

7.   V/Ω/CAP/        /Hz/%/Temp

      (positive) input jack

8.   °C/°F select button

9.   HOLD    

/Backlight button

10. Frequency/Duty Cycle

      select button
11. Relative button
12. Range select button

2.

Summary of Contents for 11060S

Page 1: ...Contents Made in China Product distributed by Southwire Company LLC One Southwire Drive Carrollton GA 30119 2014 Southwire Company LLC All rights reserved southwiretools com 1 855 SW T00LS Toll Free...

Page 2: ...be sure to secure the back cover panel and battery compartment door rmly to maintain the waterproof and dust proof integrity of the meter Loose or overtightened screws or an improperly seated o ring...

Page 3: ...ary distribution panels Overhead or underground lines to detached buildings Incoming service entrance from utility Outdoor pumps WARNING Operation is limited to CAT II applications when the insulated...

Page 4: ...red value 4 Press the REL button to return to normal operation HZ Button The HZ button is used to select frequency or duty cycle in when in the frequency voltage or current function When in the voltag...

Page 5: ...urrent AC or DC will appear on the LCD display AC and DC Current Measurements Observe all safety precautions when working on live voltages WARNING Observe all safety precautions when working on live v...

Page 6: ...ible tone will sound If the circuit is above 400 the LCD display will indicate OL The resistance will be shown on the LCD display if it is below 400 Wire 10 WARNING Operation 1 Set the rotary function...

Page 7: ...h a fresh 9V battery 5 Install battery door and tighten screws WARNING TO AVOID ELECTRICAL SHOCK REMOVE ALL INPUTS BEFORE OPENING THE CASE TO PREVENT FIRE INSTALL FUSE WITH AMP VOLT RATING SHOWN 9V BA...

Page 8: ...Fuse DC Current Auto ranging 400 0uA 4000uA 40 00mA 400 0mA 10A 0 1uA 1uA 10uA 100uA 10mA 1 0 of rdg 5 digits 2 5 of rdg 3 digits Maximum Input 4000uA DC on uA range 400mA DC on mA range 10A DC on 10A...

Page 9: ...tenance or repair operation outside of specifications or abnormal handling of the product Southwire s sole liability and the purchaser s exclusive remedy for any breach of this warranty is expressly l...

Reviews: