The OMD-625 oxygen analyzer gives the users the flexible to select a full scale range to better match
the application they are working with. For example these can be basic such as 0 - 200 ppm or can
be a bit more exotic such as 0 - 93.2 ppm.
To configure a custom range option, enter the MENU --> SYSTEM --> CUSTOM RANGE. From here,
the user can use the UP or DOWN arrows to customize the upper end of the full scale range such as
0 - 200 ppm or 0 - 93.2 ppm. Once the value is correct hit ENTER to store the custom range.
Once the value is stored, the user will now need to select the manual range option 6 in the MANUAL
RANGE mode to lock the analog output to the new custom range. This can be done by going to
MENU --> MANUAL RANGE --> And then selecting the 6th range which should match the value you
just entered.
Once selected the HOME screen should show the custom range and have a RANGE LOCKED text right
above it.
At anytime the user can go back into the MENU and change this range.