Quick-Check PowerSaver Manifold 10 & 20 Port South-Tek
Systems
Revision: 0
Revision Date: 8/30/17
Page 25 of 29
8.2
MAINTENANCE SCREEN
8.2.1
Maintenance Test
Maintenance Test is a feature on the Quick-Check
PowerSaver Manifold which allows user to view a quick
operation of the sampling function in the unit. To utilize
Maintenance Test ensure the ON/OFF selector switch
on the Home screen is in the ON postion then navigate
to the Maintenance screen through the main Menu
page.
The Maintenance Test can be started by pressing the
button on the right, it will cycle through each of the
zones for twenty seconds each. Once the last zone is
sampled, the unit will return to normal operation.
Note: Do not turn off the Maintenance Test once it has
started, if the test needs to be stopped at any point
press the selector switch on the Home screen to the
OFF position
This test mode will take purity and flow readings on each zone although normal run purity may differ from this
reading.
8.2.2
Valve Test Mode
Valve Test Mode is a feature on the Quick-Check PowerSaver Manifold which allows the user to actuate each
individual solenoid valve to ensure proper function. To utilize the Valve Test Mode turn the ON/OFF selector switch
on the Home screen to the OFF postion then navigate to the Maintenance screen through the main Menu page.
The Valve Test Mode can be enabled or disabled by toggling the button on the right. Once enabled, each valve on
the page may be actuated by pressing the corresponding button. This mode can be used specifically for
troubleshooting purposes or pressure testing.
Figure 25: Maintenance Screen
Figure 26: Valve Test Mode