27/04/04 ©
Söring
Users manual RF-units
1
not valid for MBC600
-33-
Mode
Power output into
[
Ω
]
Output range
[W]
Nominal
Frequency [kHz]
Output Volt.
[V
PP
]*
Force 1
5W [200
Ω
]
1W 30W
**
860
Force 2
30W [200
Ω
]
5W 55W
**
1150
Force 3
55W [200
Ω
]
30W 80W
**
1300
Force 4
80W [200
Ω
]
55W 120W
**
1320
Force 5
120W [200
Ω
]
80W 120W
**
1320
Spray 1
20W [200
Ω
]
20W 40W
**
9000/4800
Spray 2
40W [200
Ω
]
20W 65W
**
9000/4800
Spray 3
65W [200
Ω
]
40W 90W
**
9000/4700
Spray 4
90W [200
Ω
]
65W 120W
**
9000/5800
Spray 5
120W [200
Ω
]
90W 120W
**
9000/5800
Bipo. Cut
1
1 - 130W [200
Ω
]
-
-
360
330 - 760
Bipo. Coag
1 - 140W [100
Ω
]
-
-
360
160 - 460
Ignition/Working voltage
*) Output voltages are no load values
**) pulse shaped signal
Signal shapes
Cut
1-5 (step) continuous sine wave
Blend
1-5 (step) pulsed sine wave, duty cycle 80%
TUR
1-5 (step) continuous sine wave
Pulse
1-5 (step) continuous sine wave,
pulsed
Soft
1-5 (step) continuous sine wave
Force
1-5 (step) pulsed sine wave, duty circle 25%
Spray
1-5 (step) spike shaped signal
Bipolar Cut
1
1-130 (watt) continuous sine wave
Bipolar Coag
1-140 (watt) continuous sine wave
Bipolar Pulse
1
1-130 (watt) continuous sine wave
Visual and acoustical signals
Activation
RF on
Visual signal
Acoustical signal
Power on
Green switch light
Cutting
X
Green working indicator
Continuous tone 440Hz
Coagulation
X
Green working indicator
Continuous tone 480Hz
Bipolar cutting
1
X
Green working indicator
Continuous tone 550 Hz
Bipolar Coagulation
X
Green working indicator
Continuous tone 550Hz
patient
electrode
error
Red blinking signal
Modulated warning tone 370Hz
dosage error
X
Red signal
The level of the acoustical signal is adjustable between 45 and 65 dB (A). The level of the warning tone
(Patient electrode failure) is fixed at 65 dB (A).
Autostop time
Auto:
1.5sec.