4
Important EMC notices for use in the medical environments
• The UP-991AD/UP-971AD needs special precautions
regarding EMC and needs to be installed and put into
service according to the EMC information provided in this
instructions for use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the UP-991AD/UP-
971AD.
Warning
The use of accessories and cables other than those
specified, with the exception of replacement parts sold by
Sony Corporation, may result in increased emissions or
decreased immunity of the UP-991AD/UP-971AD.
Warning
If the UP-991AD/UP-971AD should be used adjacent to or
stacked with other equipment, it should be observed to
verify normal operation in the configuration in which it will be
used.
Guidance and manufacturer’s declaration-electromagnetic emissions
The UP-991AD/UP-971AD is intended for use in the electromagnetic environment specified below.
The customer or the user of the UP-991AD/UP-971AD should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The UP-991AD/UP-971AD uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The UP-991AD/UP-971AD is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies