4-8. Log Output
The failure analysis log can be outputted.
Procedure:
1. Enter the test mode.
(Refer to “SETTING THE TEST MODE” on page 22)
2. Press the [
v
] or [
V
] keys to select the “Log Output”, and press
the [
N
] key.
3. Press
the
[
N
] key again, and the log fi le is output to the root
folder of this unit.
<Log Output>
Success
LCD Display
4. Release the test mode, connect with PC and check the log fi le.
Root folder of this unit
(Display of PC)
PC
to a USB port
USB connecting
cable (supplied)
to the
connector
Date of log outputted
Releasing method:
Press the [
x
STOP] key, return to test mode menu.
4-9. Menu Init
The menu settings of this unit can be return to the shipment set-
tings.
Procedure:
1. Enter the test mode.
(Refer to “SETTING THE TEST MODE” on page 22)
2. Press the [
v
] or [
V
] keys to select the “Menu Init”, and press
the [
N
] key.
3. Press the [
N
] key again, and the initialization is performed.
<Menu Init>
LCD Display
Success
Releasing method:
Press the [
x
STOP] key, return to test mode menu.
4-10. Factory
Ship
This unit can return to the factory shipment settings (LCD contrast
is not initialized).
Procedure:
1. Enter the test mode.
(Refer to “SETTING THE TEST MODE” on page 22)
2. Press
the
[
v
] or [
V
] keys to select the “Factory Ship”, and press
the [
N
] key.
3. Press the [
N
] key again, and the initialization is performed.
<Factory Ship>
Execute...
LCD Display
4. This unit is reboots.