22
10.
AUTO (automatic exposure) button
11.
BLUE (B-gain) knob
12.
RED (R-gain) knob
Warning
Using this unit for medical
purposes
The connectors on this unit are not isolated.
Do not connect any device other than one
which conforms to IEC 60601-1 standards.
When an information technology device or
AV device that uses an alternating current is
connected, current leakage may result in an
electric shock to the patient or operator.
If use of such a device is unavoidable,
isolate its power supply by connecting an
isolation transformer, or by connecting an
isolator between the connecting cables.
After implementing these measures,
confirm that the reduced risk now conforms
to IEC 60601-1 standards.
Caution
Do not come into contact with the terminals
of the rear panel connectors and patients at
the same time.
Doing so may result in a generation of
voltage that can be harmful to patients if the
unit is malfunctioning.
Always disconnect the power cord before
connecting and disconnecting connectors.
1.
2.
3D-SYNC IN (3D-SYNC input)
connector,
3D-SYNC OUT (3D-
SYNC output) connector (BNC type)
3.
VIDEO (composite video) output
connector (BNC type)
4.
Remote contact switch connector 1, 2
(stereo mini jack)
5.
Equipotential ground connector
Used to make an equipotential ground
connection.
6.
-
(power) connector
7.
RS-232C connector (D-sub 9-pin)
8.
S VIDEO output connector (mini
DIN 4-pin)
Camera control unit (CCU) rear panel