Important EMC notices for use in medical environments
The LMD-XH550MT needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in the instructions for use.
The LMD-XH550MT is intended for use in a professional healthcare facility environment.
Portable and mobile RF communications equipment, such as cellular phones, can affect the LMD-
XH550MT.
Warning
Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of
the LMD-XH550MT. Otherwise, degradation of the performance of this equipment could result.
If the LMD-XH550MT will be used adjacent to or stacked with other equipment, normal operation of the
LMD-XH550MT under such configurations should be verified via observation.
The use of accessories and cables other than those specified, with the exception of replacement parts
sold by Sony Corporation, may result in increased emissions or decreased immunity of the LMD-
XH550MT.
Guidance and manufacturers declaration electromagnetic emissions
The LMD-XH550MT is intended for use in the electromagnetic environment specified below. The
customer or the user of the LMD-XH550MT should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The LMD-XH550MT uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
CISPR 32
Class B
The LMD-XH550MT is suitable for use in all
establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies