Auto Delete Settings
Select whether to use the auto delete function, and
configure the threshold value that will be used for the
function.
1
Select [Auto Delete] in the [System Admin Settings]
screen, and press the ENTER button.
The [Auto Delete Setting] screen appears.
2
Configure each setting, select [Apply] and press the
ENTER button.
[Auto Delete]
Select whether to use the auto delete function.
The default setting is [Use].
[Auto Delete Value]
Select the threshold value that will be used for the
auto delete function.
You can select from [10%], [20%], and [30%].
The default setting is [10%].
The configuration is saved.
DICOM Settings
Configure settings that are necessary for DICOM
transmissions of still images.
1
Select [DICOM Settings] in the [System Admin
Settings] screen, and press the ENTER button.
The [DICOM Settings] screen appears.
2
Select a tab, and press the ENTER button.
3
Configure the settings in each tab as necessary.
For details on each tab, see the following.
[Local Station 1] Tab (page 80)
[Local Station 2] Tab (page 80)
4
When you finish configuration, select [Apply] and
press the ENTER button.
The configurations are saved, and the [System Admin
Settings] screen appears again.
[MWM Server 1] Tab
Configure MWM server settings.
[MWM]
Select whether to use the MWM server.
The default setting is [Not Used].
[Confirm] appears when [Use] is selected. Confirm
communication with the MWM server. For details, see
“To confirm communication with the MWM server”
(page 79)
[Server Name]
Enter the server name.
This item is available when [MWM] is set to [Use].
Up to 253 characters can be entered.
[Port No.]
Enter the port number.
This item is available when [MWM] is set to [Use].
You can enter a value ranging from 1 to 65535 (up to 5