15
Installation guide | Supercal 5
EN
Communication options
The Supercal 5 can be fitted with up to two different optional communication
modules. The optional communication modules can be equipped afterwards,
without damaging the verification. The optional modules have no influence
on the verified relevant part in the cover of the inte-grator unit. At the latest
6 seconds after the installation, the calculator unit recognizes the plugged-in
optional modules and the functions are freely available. When connecting the
communication modules, the installation guidance - supplied with the unit - is
to be considered.
Cooling liquids (Glycols)
In the calculator Supercal 5 more than 70 cooling liquids are programmed and
many additional mixtures can be specified by software.
The feature of the calculator Supercal 5 for cooling applications with cooling
liquids water mixtures is exclusively to be used with the flow sensor Superstatic
440 (Not to be used with mechanical flow sensors).
The feature cooling liquid (Glycol) water mixture is enabled with the Superprog
windows and Superprog android.
Note: If cooling liquids are used, the calculator or thermal energy meter loses
its MID approval.
Display
The calculator Supercal 5 has the following display sequence:
Main menu (Billing relevant data)
Metrological
Configuration
Service
Test Radio