24
Chapter 4:
Replacement Procedures
Power Supply PCBA Replacement
Power Supply
Removal
1
Gently pry the shield from the power supply and set it aside. This part will be used in
reassembly. Note that the shield fits only one way. See
Figure 4.9 Power Supply Shield
2
Remove the 7 screws that hold down the power supply PCB per
.
3
Gently lift the power supply away from the Main PCBA.
4
Install the new Power Suppply PCBA by reversing steps 1-3.
Figure 4.10 Power Supply Screws
Power supply
shield
Power supply
screws (7x)