Page
28
of
29
Step
Description
Reference
4
Attach a clean, dry, regulated compressed air
source to one of the ¼” Air Cooling hose barbs.
Note: The compressed air should be oil free and
filtered with a 5 micron filter.
5
A 2
nd
hose must be attached to the 2
nd
fitting.
This hose will exhaust the air from the converter
outside of the enclosure.
6
The exhaust hose only needs to exit the top of the
enclosure by at approximately 8 inches or more.
COOLING AIR REGULATION AND ADJUSTMENT
Adjust the regulated compressed air until the gauge indicates 10 ± 1 PSIG (approximately 3 to 4 CFM).
The compressed air will flow into the converter and out through the outlet barb.
Verify the compressed
air is flowing out from the exhaust tube upon each use.