Page
15
of
29
7.2.
SETUP
1. Connect the power cord into the receptacle on the rear of the ultrasonic processor.
2. Make sure the unit is switched off. Plug the electrical line cord into the electrical outlet.
3. If the optional foot switch is used, insert the plug into the jack located on the rear panel.
4.
For best results, it is critical to use the appropriate size and type of accessory to
process your sample.
If you are not sure that you have the proper probe for your sample
volume, please refer to the product brochure or call for assistance.
5.
Horns/Probes must be properly tightened.
Depending on the accessories purchased,
often the probe and the flat tip are attached to the converter at the factory.
Check the
tightness of the probe and flat tip by using the wrench set.
Please refer to images in
the Maintenance section of this manual. A loose probe or tip may cause damage to the
power supply circuitry or parts of the converter and probe. A loose probe may also show a
fluctuation in wattage readings.
Always use the wrenches supplied with the unit.
6. If you will be using a microtip or extender, remove the flat tip on the end of the replaceable
tip probe, then attach the microtip or extender in its place.
7.
Microtips must be used in pulse mode to prevent overheating
- which could potentially
crack the tip. See page 16 for Microtip Limits and contact us with questions.
8. Horns and probe tips wear after normal usage. Using a severely worn probe tip can
damage internal power supply components.
9. If using a laboratory stand, mount the converter /probe assembly using a clamp. Be sure to
secure the clamp to the upper section of the converter housing only. Never secure the
clamp to any other portion of the converter/probe assembly. If you are using an acoustic
enclosure mount the converter properly in the converter collar.
10. Connect the converter cable to the power supply and then to the top of the converter. Push
the connectors in and turn the chrome rings clockwise ¼ turn to secure the connectors.
11. If the application requires long processing times, we recommended chilling the sample and
pulsing processing. If processing for over 15 minutes, the converter may get warm and
require cooling with dry, compressed air.
See converter cooling instructions in the
Section 12.