Measurement limit Resolution
Accuracy
200 µA
0.1 µA
±(1.2% 3 digits)
2 mA
1 µA
20 mA
10 µA
200 mA
100 µA
10 A
10 mA
±(2.0% 5 digits)
Overload protection: an F250 mA / 600 V fuse at the
VΩ Hz
mAμA
socket and an F10 A/600 V fuse at the “10A”
socket. Max input current:
VΩ Hz
mAμA
socket: 250 mA DC
or AC RMS; at “10A” socket: 10 A DC or AC RMS.
If the measured current is greater than 5 A, the dura
-
tion of continuous measurement should not exceed
10 seconds, and at least 5 minute should elapse be
-
tween two measurements.
Measurement of AC voltage (
V
)
Connect the red probe cable to the
VΩ Hz
mAμA
socket,
and the black one to the “COM” socket.
Set the dial to the AC voltage position. Connect the
measurement probes to the circuit to be measured.
Note: The reading can be unstable, especially in the
200 mV measurement range, if the probe cable is not
connected to the circuit being measured.
Measurement limit
Resolution
Accuracy
2 V
1 mV
±(1% 3 digits)
20 V
10 mV
200 V
100 mV
600 V
1 V
±(1.2% 3 digits)
Overload protection: Use of air-gap (for voltages over
1500 V)
Measuring range: 40 Hz- 200 Hz
Measurement of AC (
μ
A mA A
)
Connect the black probe cable to the “COM” socket.
Connect the red probe cable and set the selector dial
according to the current values to be measured:
Current to be
measured
Red probe cable
connection (socket)
Dial setting
0 – 2 mA
VΩ Hz
mAμA
μ
A
2 – 200 mA
VΩ Hz
mAμA
mA
0.2 – 10 A
10 A
A
Use the “FUNC” pushbutton to set the AC (
AC
)
function. Interrupt the circuit whose current is to be
measured then connect the probe to the points to be
measured.
Measurement limit
Resolution
Accuracy
200 µA
0.1 µA
±(2.5% 5 digits)
2 mA
1 µA
20 mA
10 µA
200 mA
100 µA
10 A
10 mA
Overload protection: an F250 mA / 600 V fuse at the
VΩ Hz
mAμA
socket and an F10 A/600 V fuse at the “10A”
socket. Max input current:
VΩ Hz
mAμA
socket: 250 mA DC
or AC RMS; at “10A” socket: 10 A DC or AC RMS. If
the measured current is greater than 5 A, the dura
-
tion of continuous measurement should not exceed
10 seconds, and at least 5 minute should elapse be
-
tween two measurements.
Measuring range: 40 Hz – 400 Hz
Frequency measurement
Connect the red probe cable to the
VΩ Hz
mAμA
socket,
and the black one to the “COM” socket. Set the dial to
the “Hz” function. Connect the measurement probe to
the circuit to be measured.
Measurement limit
Resolution
Accuracy
20 kHZ
10 Hz
±(1.5% 5 digits)
Input signal level: min. 500 mV
Diode and interruption testing
Diode testing: Connect the red probe cable to the
VΩ Hz
mAμA
socket, and the black one to the “COM” socket
(the red probe is for “+” polarity). Set the dial to the
position. Use the “FUNC” pushbutton to set the
diode testing (
) function. Connect the red probe
cable to diode’s anode, and the black one to its cath
-
ode. The display will show the diode’s approximate
forward voltage. If the connection is reversed, “OL”
will appear in the display. Interruption testing:
Con-
nect the red probe cable to the
VΩ Hz
mAμA
socket, and
the black one to the “COM” socket. Set the dial to the
position. Use the “FUNC” pushbutton to set the
interruption testing (
) function. Connect the mea
-
surement probe to the circuit to be tested. If the circuit
is connected to a power source, switch of the power
source and discharge the capacitors before starting
measurement. Short-circuits (less than 100 Ω) will be
indicated by the built-in buzzer.
Function
Resolution
Measuring environment
1 mV
Measurement current: approx. 1
mA Idle voltage: approx. 1.48 V
The built-in buzzer will
activate below 100 Ω
Idle voltage less than 500 mV
5