Page 28
Transcend Sleep Apnea Therapy User Manual
Caring for your Transcend and
components
This section presents the following topics:
•
Cleaning the exterior
•
Cleaning of accessories
•
Cleaning the filter media and filter frame
Warning:
•
Unplug the Transcend before cleaning.
•
Do not submerge the Transcend or power supply in liquid.
•
Prevent water from entering any openings of the device.
•
Do not use harsh or abrasive cleaning agents to clean the device or any
components.
•
Do not attempt to sterilize the Transcend.
•
Do not place cleaning materials, such as a cloth or liquid, into the device air inlet or
air outlet connector.
Cleaning the Exterior
Follow these instructions to clean the exterior of the Transcend device.
1.
Unplug the power supply prior to cleaning and disconnect the device from power cords.
2.
Mix a solution of 5% mild liquid detergent in distilled water (1.6 fl oz liquid detergent
per quart of distilled water). Mild detergent should contain biodegradable anionic
surfactants and no phosphate.
3.
Submerge a lint-free cotton cloth into the detergent solution.
4.
Wring excess water from the cloth then wipe the exterior of the Transcend device for
approximately 20 seconds using a gentle, back and forth wiping motion from the front
to back of the device. Apply firm pressure and ensure contact with all accessible
contact surfaces to adequately remove soil buildup.
5.
Rinse the cloth in clear water to remove residual cleaning solution.
6.
Wring excess water from the cloth then wipe the Transcend device using a gentle front
to back wiping motion to remove any detergent solution remaining on its surface.
7.
Wipe the device with a dry, lint-free cotton cloth until the device is fully dry.